KP-DFT201 Defibrillation Effect and Energy Tester Delivered to DEKRA Guangzhou Laboratory

Table of Contents

Medical Electrical Safety · Technical Delivery Case

KP-DFT201 Defibrillation Effect and Energy Tester Delivered to DEKRA Guangzhou Laboratory

KingPo delivered a KP-DFT201 Defibrillation Effect and Energy Tester to DEKRA Guangzhou Laboratory for medical electrical safety testing. The delivered system supports laboratory work involving protection against defibrillation effects, energy-related measurement, residual-voltage assessment and oscilloscope waveform observation for patient-connected medical electrical equipment.
The laboratory setup combines the KP-DFT201 with an external oscilloscope. Test conditions are configured through the tester’s touchscreen, while the oscilloscope provides an independent observation path for the applied waveform and connected measurement channels. This configuration enables the laboratory to review the complete pulse and measurement chain rather than relying on the high-voltage source alone.
The front panel integrates the high-voltage output, return connection, selectable output terminals and dedicated measurement points required for the test setup. Manual and automatic operating selections are visible on the touchscreen, together with standard selection, test item, voltage, polarity, pulse count, interval and connection settings.
KP-DFT201 defibrillation effect and energy tester connected to an oscilloscope in a medical electrical safety laboratory
Illustrative laboratory image based on the delivered KP-DFT201 setup, showing the tester connected to an external oscilloscope for waveform observation.

Project Overview

Customer
DEKRA Guangzhou Laboratory
Delivered Model
KP-DFT201
Main Application
Defibrillation effect and energy testing
Measurement Setup
Tester with external oscilloscope
Primary Test Functions Protection against defibrillation effects, energy-related measurement, residual-voltage-related assessment and waveform observation
Operating Interface Touchscreen control with manual and automatic test selections
External Instrument External oscilloscope connected to the tester’s measurement terminals
Visible Standard Selection GB 9706.1-2020 / IEC 60601-1:2020 is displayed on the photographed touchscreen
Project Evidence Delivered equipment photographs, front-panel connection details and oscilloscope waveform images

Delivery Context and Laboratory Objective

The project was configured for a professional third-party medical electrical safety laboratory. In this environment, the tester must support repeatable setup, clearly identified connection points, controlled pulse sequencing and a measurement arrangement that can be reviewed by different laboratory personnel.
A laboratory test system also needs to fit into an existing quality workflow. The operator must be able to record the selected standard, circuit, polarity, test voltage, pulse count, interval, oscilloscope configuration and post-test observations. These records are necessary for repeatability and technical review, even when the tester provides an automatic sequence.
For similar projects, KingPo reviews the medical equipment type, applied parts, patient connections, required common-mode or differential-mode tests, energy-reduction requirements, oscilloscope configuration and calibration scope before confirming the final tester and accessory arrangement.

Laboratory Testing Requirement

Medical electrical equipment connected to a patient may be exposed to defibrillation pulses during clinical use. A laboratory evaluating this type of equipment therefore requires a controlled pulse source and measurement arrangement capable of applying the specified condition while monitoring the response of the equipment under test.
The test cannot be defined only by peak voltage. The applicable circuit, waveform, resistance, inductance, polarity, pulse count, interval, patient connection, operating condition and post-test assessment must all correspond to the relevant standard clause and type of medical electrical equipment.
The selected system needed to support both test execution and waveform review. The KP-DFT201 provides the configurable pulse and connection interface, while the oscilloscope allows the laboratory to observe signals at designated measurement points and confirm that the intended test arrangement has been established.
The delivered configuration is therefore more than a standalone high-voltage source. It is a coordinated test platform involving the pulse source, selected circuit, connection leads, measurement terminals, oscilloscope channels, attenuation arrangement and equipment under test.

How the Defibrillation Effect and Energy Tester Supports Laboratory Testing

Protection against defibrillation effects

This test evaluates the response of patient-connected medical electrical equipment when the specified defibrillation pulse is applied. The selected test path, waveform, polarity, pulse sequence, operating condition and post-pulse evaluation must follow the applicable standard requirements.
Energy reduction and energy-related measurement

The energy-related function is used to evaluate energy behaviour through the configured circuit. Resistance, inductance, waveform, voltage and measurement path must match the required test arrangement rather than being selected only from the maximum capability of the instrument.
Residual-voltage assessment

Where required by the applicable standard, the laboratory may need to evaluate residual or hazardous voltage after the pulse. The measurement point, time window, circuit condition and applicable limit must be established before testing.
Functional recovery and essential performance

The equipment under test may also require functional or essential-performance checks after exposure. A compliant pulse waveform alone does not demonstrate that the medical equipment has met all post-test requirements.

Delivered Equipment and Front-Panel Configuration

The delivered instrument is identified as KP-DFT201. Its front panel integrates the main output and measurement connections required for a controlled defibrillation test setup. The layout separates the high-voltage output from the return path, selectable output terminals and oscilloscope measurement connections.
The touchscreen provides access to the test mode, standard selection, test item, waveform, voltage, polarity, pulse count, interval and connection state. This reduces the need to configure separate external switching devices for every routine sequence, although the operator must still verify the actual wiring before enabling the high-voltage output.
KingPo’s current Defibrillation Effect and Energy Tester series can be reviewed for related configurations. The linked page is a product-family reference; the equipment shown in this delivery case is specifically marked KP-DFT201.
Visible Interface Role in the Laboratory Setup
HV OUT Dedicated high-voltage output used for the configured defibrillation pulse path
RETURN / Ground Return connection used to complete the selected test circuit
50 Ω Output Selectable output terminal used when the applicable circuit requires the corresponding resistance path
400 Ω Output Alternative output terminal for the relevant configured test path
AUX Output Auxiliary connection used according to the selected test arrangement
MEAS. Terminals Dedicated measurement points for oscilloscope connection and channel comparison
Touchscreen Interface Configuration of the selected standard, test item, waveform, voltage, polarity, pulse sequence and connection status
Close-up of the KP-DFT201 front panel with touchscreen, high-voltage output and measurement terminals
Illustrative close-up based on the delivered equipment, showing the touchscreen, return terminal, selectable outputs, high-voltage output and measurement connections.

What This Tester Is—and What It Is Not

The KP-DFT201 is used to apply simulated defibrillation-related test conditions to patient-connected medical electrical equipment and to support protection, energy-reduction and residual-voltage test workflows.
It should not be confused with a defibrillator analyser used to measure the delivered energy, current, voltage and biphasic waveform of an actual clinical defibrillator or AED under IEC 60601-2-4-related output-performance tests. These are different test objectives and normally require different measurement systems.
Selection mistake to avoid

A request described only as “defibrillation testing” is not technically sufficient. KingPo must first confirm whether the customer needs to test a defibrillator’s output performance, a patient-connected device’s immunity to defibrillation pulses, defibrillation-electrode performance, or an IEC 60601-2-31 test pulse.

Oscilloscope and Waveform Observation Setup

The delivered setup uses an external oscilloscope connected to the KP-DFT201 measurement terminals. Two visible oscilloscope channels allow the laboratory to observe the configured signal and compare the response captured through the selected measurement path.
The oscilloscope is not merely a display accessory. Probe attenuation, input range, vertical scaling, bandwidth, coupling, grounding, trigger source, trigger level and the exact location of each measurement point can all affect the displayed waveform.
Before a trace is accepted as representative, the laboratory should confirm that the oscilloscope probes and measurement channels are suitable for the voltage, frequency content and transient behaviour of the test. The attenuation factor entered into the oscilloscope must correspond to the actual probe or external attenuation network.
Waveform observation is not a pass/fail result

A photograph of a waveform confirms that a signal was displayed during the setup. It does not by itself establish the calibrated amplitude, energy value, measurement uncertainty, applicable limit, functional recovery or final conformity decision.
Oscilloscope waveform observation setup connected to the KP-DFT201 defibrillation effect tester
Illustrative setup based on the delivered system, showing an external oscilloscope used to observe the configured pulse and measurement-channel response.

Test Parameters That Must Be Controlled

The operator should configure and record the complete test condition. Using the correct voltage with an incorrect circuit or measurement path can still produce an invalid test.
Parameter Why It Matters
Applicable standard and clause Defines the circuit, pulse condition, specimen state, measurement method and acceptance criteria
Common-mode or differential-mode path Determines how the pulse is applied to the patient connections and reference points
Waveform and voltage Defines the electrical stress applied to the equipment under test
Polarity Some requirements call for positive, negative or both pulse polarities
Resistance and inductance Shape the pulse and determine the energy delivered through the selected test circuit
Pulse count and interval Controls the complete exposure sequence and recovery time between pulses
Equipment operating condition Determines whether normal function, monitoring or essential performance must be maintained during and after exposure
Post-test evaluation Defines residual-voltage, recovery, functional or essential-performance checks required after the pulse sequence

Why the Complete Measurement Chain Matters

Defibrillation testing involves more than the internal pulse source. The tester, selected output path, circuit components, high-voltage leads, return connection, oscilloscope, probes, attenuation devices, grounding arrangement and equipment under test all influence the measured result.
Test-circuit selection

The laboratory must identify the applicable applied parts, patient connections, test terminals and pulse path before wiring the equipment. A circuit selected for one medical device cannot automatically be transferred to another device with a different patient connection arrangement.
Cable and lead routing

High-voltage output leads and oscilloscope measurement cables should be routed consistently and kept away from unrelated signal cables, accessible conductive parts and operator contact areas. Lead position can influence coupling, noise and repeatability.
Oscilloscope channel assignment

Each measurement point should be linked to a clearly identified oscilloscope channel. Channel labels, attenuation ratios, vertical scales and time-base settings should be recorded in the test sheet so that the displayed traces can be interpreted correctly.
Post-pulse evaluation

After the pulse sequence, the laboratory may need to assess residual voltage, operational recovery, alarms, displayed values, essential performance or other product-specific functions. The required checks should be defined before the pulse is applied.

Recommended Laboratory Test Workflow

A controlled workflow helps reduce wiring errors and prevents a visually acceptable waveform from being mistaken for a complete conformity result.
  1. Confirm the applicable IEC 60601 or GB 9706 standard, edition, clause and particular-standard requirements.
  2. Identify the type of medical equipment, applied parts, patient connections, accessible conductive parts and protective-earth arrangement.
  3. Determine the required pulse circuit, waveform, polarity, resistance, inductance, voltage, pulse count and interval.
  4. Connect the high-voltage output, return path and equipment under test according to the approved wiring diagram.
  5. Connect the oscilloscope to the designated measurement points and record the probe or attenuation ratios.
  6. Confirm channel scaling, coupling, bandwidth, trigger source, trigger level and time base before enabling the pulse output.
  7. Place the equipment under test in the operating state required by the applicable standard.
  8. Run the required pulse sequence while maintaining safe clearance from high-voltage connections.
  9. Record the waveform, measured values, pulse condition and any observed behaviour of the equipment under test.
  10. Complete the specified residual-voltage, recovery, functional, dielectric or essential-performance assessment after exposure.
Three-quarter laboratory view of the KP-DFT201 defibrillation tester and oscilloscope
Illustrative test-bench view based on the delivered setup. The tester and oscilloscope provide a coordinated platform for pulse configuration, waveform observation and measurement review.

Applicable Standards and Test Boundaries

The touchscreen shown in the actual delivery photograph includes a GB 9706.1-2020 / IEC 60601-1:2020 selection. This confirms that the interface contains a standard-related configuration menu, but it does not mean that one test circuit or acceptance criterion applies to every medical electrical device.
General medical electrical safety requirements may be supplemented by particular standards for electrocardiographs, patient monitors or other patient-connected equipment. Particular standards can modify the applied-part configuration, pulse sequence, test circuit, operating state, recovery period and acceptance criteria.
Laboratories planning broader capability can review KingPo’s IEC 60601-1 test equipment list and range of medical electrical safety test equipment for related leakage-current, electrosurgical, ECG, defibrillation and other medical-device test applications.
Standard Context Application Boundary
IEC 60601-1 / GB 9706.1 General medical electrical safety requirements, including protection against defibrillation effects where applicable
ECG-related particular standards Additional requirements for electrocardiographs, patient monitors, ECG monitoring equipment and associated patient connections
Other particular standards Product-specific requirements may define different circuits, operating conditions, pulse sequences, recovery times and acceptance limits
Do not configure the test from voltage alone

A valid defibrillation-effect test requires the specified waveform, circuit impedance, polarity, pulse sequence, connection mode, specimen operating state, measurement method and post-test assessment. Matching only the peak voltage is insufficient.

Laboratory Preparation and Safety Considerations

The KP-DFT201 is used in a high-voltage transient test environment. The laboratory should establish a controlled operating area, approved wiring diagrams and a repeatable connection procedure before routine testing begins.
Controlled test area

The tester, oscilloscope and equipment under test should be positioned so that operators can view the interfaces without reaching across energized leads. Unrelated personnel should be kept outside the test area during pulse application.
Connection verification

Output, return and measurement leads should be checked against the approved test diagram before the high-voltage output is enabled. Connections should not be changed while stored energy may remain in the circuit.
Grounding and measurement isolation

The grounding arrangement of the tester, oscilloscope and equipment under test must be reviewed as a complete system. An incorrect ground connection can change the intended pulse path or create an unsafe condition.
Discharge and waiting procedure

The laboratory should define how stored energy is discharged and how the operator confirms a safe state before disconnecting the equipment under test or moving measurement leads.

Documentation, Verification and Calibration Planning

Before the tester enters routine use, the laboratory should define the documentation required by its quality system. Relevant records may include operating instructions, wiring diagrams, approved test methods, equipment identification, connection photographs, oscilloscope setup records, maintenance instructions and software or configuration records.
Verification and calibration should be planned by measurement parameter rather than by the product name alone. The tester and oscilloscope form part of one measurement chain, and each relevant parameter should be linked to the laboratory’s traceability and recalibration plan.
Depending on the applicable method, the scope may need to include:
  • High-voltage output magnitude and polarity
  • Pulse waveform and timing characteristics
  • Resistance and inductance values in the selected test circuit
  • Pulse count and interval timing
  • Energy measurement or energy-calculation method
  • Oscilloscope input scaling and external attenuation ratio
  • Measurement-channel bandwidth and timing accuracy
  • Residual-voltage measurement function where applicable
  • Software, test-sequence and connection-state verification where required
Project evidence boundary

The supplied photographs confirm the delivered model, front-panel configuration, laboratory setup and waveform-observation arrangement. They do not by themselves establish calibration status, measurement uncertainty, formal customer acceptance or a test pass/fail result.

Practical Value for a Medical Electrical Safety Laboratory

A touchscreen-controlled system can reduce repetitive manual switching and improve consistency between test sequences. However, the value of automation depends on correct circuit selection, verified wiring and a documented measurement process.
Dedicated output and measurement terminals also make the setup easier to inspect. Operators can identify the high-voltage path, return path and oscilloscope connections without relying on temporary external wiring for every routine test.
For laboratories testing different categories of patient-connected equipment, the main advantage is configurability. The tester can be reviewed against the laboratory’s applicable standards, test circuits, lead arrangements and documentation requirements instead of being selected only by a headline voltage specification.

Engineering Notes for Similar Laboratories

Confirm the exact standard and clause

The phrase “defibrillation test” is not enough to define a configuration. The laboratory should identify the general standard, particular standard, edition, clause, device type and patient connection arrangement before selecting the pulse circuit.
Verify the complete measurement chain

The tester, oscilloscope, probes, cables, attenuation devices, circuit components and grounding arrangement all influence the recorded waveform. Verification should cover the same configuration used during routine testing.
Separate waveform review from conformity assessment

A visible waveform is useful for confirming setup and signal behaviour, but conformity must be based on the specified measurement method, numerical limits, operating condition and post-test performance criteria.
Record every configurable parameter

Voltage, polarity, pulse count, interval, selected output, measurement channel, oscilloscope range and specimen operating mode should be recorded so that the test can be repeated and independently reviewed.
Define the post-test assessment before exposure

The laboratory should not decide after the pulse which functions to inspect. Residual voltage, recovery time, alarms, displayed values, essential performance and other required checks should be specified in the test plan before exposure.

Information Required for a Similar Test System

To configure a similar defibrillation effect and energy measurement system, the laboratory should provide:
  • Applicable IEC 60601, GB 9706 or particular-standard references and editions
  • Type of medical electrical equipment being tested
  • Applied-part classification and patient connection arrangement
  • Required common-mode and differential-mode test circuits
  • Required waveform, voltage and polarity
  • Required resistance and inductance values
  • Pulse count, pulse interval and operating sequence
  • Energy-reduction and residual-voltage requirements
  • Number of measurement channels and oscilloscope configuration
  • Manual or automatic lead-switching requirements
  • Required test reports, calibration points and certificate type
  • Laboratory power supply, bench layout and high-voltage safety requirements

Frequently Asked Questions

What is a defibrillation effect test?
It evaluates how patient-connected medical electrical equipment responds when the specified defibrillation pulse is applied and whether the equipment continues to meet the required safety and performance criteria.
What is measured during an energy measurement test?
The exact quantity and method depend on the applicable clause and test circuit. The laboratory must define the circuit components, waveform, measurement path, calculation method and required limit before testing.
Why is an external oscilloscope connected to the KP-DFT201?
The oscilloscope provides an independent observation path for the pulse waveform and measurement channels. Its probe ratio, scaling, bandwidth, grounding and trigger settings must be verified before the trace is interpreted.
Is this the same as a defibrillator analyser?
No. This system applies simulated defibrillation-related conditions to patient-connected medical equipment. A defibrillator analyser measures the output energy and waveform of an actual defibrillator or AED.
Does the same test circuit apply to every medical device?
No. The required circuit depends on the equipment type, applied parts, patient connections and the applicable general and particular standards.
Can the KP-DFT201 run an automatic pulse sequence?
Manual and automatic selections are visible on the touchscreen. The actual degree of automation depends on the selected test circuit, connection arrangement and configured sequence.
Does a waveform image confirm that the equipment passed?
No. A waveform image confirms signal observation. A conformity decision requires verified test conditions, recorded measurement values, applicable limits and the prescribed post-test assessment.
Should the oscilloscope be calibrated separately?
The oscilloscope and any probes or attenuation devices should be included in the laboratory’s measurement traceability plan. The complete measurement chain used for the test should be verified.
Are the four article images original DEKRA site photographs?
The four inserted images are illustrative images created from the actual delivered equipment photographs to present the system more clearly. They should not be treated as independent evidence of DEKRA’s facility, test results or formal acceptance.
What information is needed to configure a similar system?
Provide the applicable standards and editions, equipment type, patient connections, required pulse circuit, waveform and voltage, energy and residual-voltage requirements, automation needs, oscilloscope setup and calibration scope.

Planning a Defibrillation Effect or Energy Measurement Test?

Send KingPo the applicable IEC 60601 or GB 9706 clauses, medical equipment type, patient connections, required pulse circuits, waveform conditions, oscilloscope setup and calibration requirements. The tester and accessory configuration can then be reviewed against the laboratory’s actual workflow.

Contact KingPo

Picture of Bruce Zhang

Bruce Zhang

Bruce Zhang is the Founder and Senior Engineer of KingPo Technology Development Limited, with over 16 years of experience in environmental and safety testing technologies. As a member of SAC TC118, TC338, and TC526, he participates in national standard reviews and provides technical guidance on IEC and ISO compliance for global laboratories.

Welcome To Share This Page:

Related Products

Related News

EU Regulatory Update · Medical Devices EU MDR Harmonised Standards 2026: Testing Updates for Medical Device Manufacturers EU MDR harmonised

DENG Sha¹, ZHANG Chao², WU Shaohai¹    Corresponding author: FAN Xiang¹ Guangdong Institute of Medical Device Quality Supervision and Inspection, No.

Defibrillation electrode testing verifies whether disposable and multifunction electrode pads can maintain stable impedance, recover after a 360J shock, control

Choose a medical bed side rail strength tester for IEC 60601-2-52 by focusing on force range, accuracy, compliance, and safety features.

When selecting a defibrillator tester (also known as a defibrillator analyzer), biomedical engineers and clinical laboratories must focus on accuracy,

Comparing IEC 60601-2-34 and IEC 60601-2-25: A Technical Guide details device scope, testing, and compliance differences for medical device standards.
KINGPO’s Electrosurgical Unit analyzer ensures ESU safety and IEC 60601-2-2 compliance with precise testing, supporting patient protection and regulatory needs.
Ensure IEC 60601 compliance in 2026 with Defib-proof and Energy Reduction Tester for safer, faster medical device approval and reliable patient protection.
Scroll to Top

Get A Free Quote Now !

Contact Form
If you have any questions, please do not hesitate to contact us.