KP3301/KP3301B Defibrillation Effect & Energy TesterโProfessional Medical Surge Tester for IEC 60601 Compliance
KP3301/KP3301B Defibrillation Effect & Energy Tester professionally simulates defibrillator pulses, accurately tests ECG monitor defibrillation protection and energy reduction per IEC 60601 standards.
MOQ: 1
Price: To be quoted
Standard Packaging: 500*500*500mm
Delivery Period: 15day
Payment Method: T/T
Supply Capacity: 100
Defibrillation Effect & Energy Tester Description
The KP3301 (and advanced KP3301B) is a high-reliability defibrillation effect and energy tester designed for testing protection against defibrillation pulses in medical devices such as ECG monitors, multi-parameter patient monitors, electrocardiographs, and ECG leads. It delivers precise defibrillation simulation pulses up to ยฑ5.5kV, supports common-mode and differential-mode protection tests, energy reduction measurement, and residual/hazardous voltage (Y1-Y2) calculations. With built-in standards compliance, automatic/manual modes, and optional auto lead switching (KP3301B), this tester ensures accurate, repeatable results for design verification, type testing, and regulatory registrationโmeeting key requirements of IEC 60601-1, IEC 60601-2-25/26/27/30/34/49, GB9706 series, YY0828, and EC13/EC53.
Product Features
- High-voltage capability with main capacitor of 32ฮผF ยฑ5% rated at 8kV, using imported solid capacitors and high-voltage vacuum relays for superior durability and reliability.
- Flexible test circuits with integrated resistance (10ฮฉ/11ฮฉ auto-switch, 50ฮฉ/100ฮฉ/400ฮฉ ยฑ1%) and inductance (500ฮผH/25mH ยฑ5%) settingsโone-click selection per applicable standards.
- Comprehensive testing functions including defibrillation effect protection (common/differential mode) for ECG/respiration, energy reduction tests with direct Joule calculation, and automatic residual voltage/hazardous voltage (Y1-Y2) measurement to eliminate oscilloscope interference errors.
- Internal signal simulation for realistic testing: sine wave 10Hz (0.5~10mV, default 5mV) and respiration rate 5~60 rpm (default 27 rpm).
- User-friendly interface with 7-inch LCD touchscreen, digital voltage/polarity control (resolution 1V, accuracy ยฑ1%), and automatic/manual test modes.
- Energy test mode using 25mH inductance, displaying resolution of 1 Joule (no-load theoretical values: ~364J at 10ฮฉ, ~360J at 11ฮฉ).
- Optional automatic defibrillator lead converter box (KP3301B model) for fully automated ECG lead switchingโno manual intervention required.
- Key Highlights & Advantages
- Superior Reliability & Longevity โ Built with premium imported solid capacitors and vacuum relays, offering exceptional stability and interference resistance compared to traditional setups.
- Automated & Efficient Testing โ KP3301B enables fully automatic lead switching and testing, significantly reducing manual effort and improving throughput for high-volume production or certification labs (vs. manual switching on KP3301).
- Accurate, Interference-Free Measurements โ Built-in automatic calculation of residual/dangerous voltage (Y1-Y2) avoids errors from external oscilloscope defibrillation interference.
- One-Click Standards Compliance โ Pre-integrated test circuits for major standards (IEC 60601 series, GB9706, YY0828, EC13/EC53), simplifying setup and ensuring repeatable, regulation-ready results.
- Versatile & Professional-Grade โ Supports both manual and automatic modes with precise digital control, ideal for R&D, quality assurance, regulatory testing, and educational/training applications.
Defibrillation Effect & Energy Tester Technical
| Parameter Name | Specification (KP3301 / KP3301B) |
|---|---|
| Charging Voltage | 0 ~ ยฑ5.5 kV ยฑ1% |
| Main Capacitor | 32 ฮผF ยฑ5% (imported solid capacitor, 8 kV rating) |
| Inductance | 500 ฮผH and 25 mH ยฑ5% |
| Internal Resistance | 10 ฮฉ and 11 ฮฉ (automatically switch or set per standard) |
| Main Resistor (non-inductive) | 100 ฮฉ, 50 ฮฉ, 400 ฮฉ ยฑ1% |
| Internal Signal | Sine wave: 10 Hz, 0.5~10 mV (default 5 mV) Respiration rate: 5~60 rpm (default 27 rpm) |
| Voltage Control | Digital settings |
| Polarity Control | Positive/negative, digital settings |
| Control & Display | 7-inch LCD touchscreen |
| Voltage Resolution & Accuracy | 1 V, ยฑ1% |
| Energy Test Mode | Operating with 25 mH inductance (meets energy reduction test requirements) Display resolution: 1 Joule |
| Energy Test Reference Values | No-load: R=10 ฮฉ โ theoretical 364 J ยฑ1% R=11 ฮฉ โ theoretical 360 J ยฑ1% |
| Y1-Y2 (Residual/Hazardous Voltage) | Automatically calculated (avoids oscilloscope interference errors) |
| Applicable Standards | GB9706.1-2020, GB9706.225/226/227-2021, IEC 60601-1:2020, IEC 60601-2-25/26/27/30/34/49, YY0828-2011, EC13, EC53, etc. |
| Model Differences | KP3301: Manual defibrillator lead switching KP3301B: Includes automatic defibrillator lead converter box for fully automated ECG lead testing |
Get Free Quote of Defib-proof and Energy Reduction Tester
Reduced our defibrillation testing time by 60% โ Leading ECG OEM
Defib-proof and Energy Reduction Tester Applications
Medical Device Manufacturers
โ Defibrillation protection and energy reduction testing for ECG monitors, multi-parameter patient monitors, electrocardiographs, and ECG lead wires.
Testing & Certification Labs
โ Compliance verification for regulatory submissions (NMPA, FDA, CE marking).
Research Institutions
โ Performance evaluation and development of defibrillation-resistant medical electronics.
Teaching & Training Programs
โ Hands-on demonstration of defibrillation effect standards and testing procedures.
Defibrillation Effect & Energy Tester Detail Display
Defib-proof and Energy Reduction Tester FAQs
Related Products
Related News
Understanding the ESU Analyzer: A Comprehensive Guide In the world of medical equipment, precision and safety are paramount. One device […]
External defibrillators are performanceโcritical devices. For thirdโparty calibration and certification laboratoriesโand for manufacturersโ QA/production teamsโthe โmake or breakโ is whether […]
IP (Ingress Protection) ratings are enclosure protection standards developed by the International Electrotechnical Commission (IEC), primarily assessing the ability of […]
Shan Chao 1 , Qiang Xiaolong 2 , Liu jiming 3 , Zhang Chao 3 . (1. Heilongjiang Institute for […]