




IEC 60601-2-31 Defibrillation Pulse Tester
Designed for defibrillation protection testing of external cardiac pacemakers with internal power source. It generates standard defibrillation pulses through a controlled RCL discharge circuit and helps manufacturers and testing laboratories verify whether the pacemaker can maintain basic safety and essential performance after exposure to the required pulse sequence.
- Model: KP-900S
- Applicable standards: IEC 60601-2-31:2020 / YY 9706.231-2023
- Test object: External cardiac pacemakers with internal power source
- Test purpose: Defibrillation protection testing and pulse withstand verification
- Pulse circuit: RCL discharge circuit
- Maximum pulse voltage: 900 V
- Energy range: 50 J to 55 J
- Built-in test network: Three-chamber external pacemaker test network
- Control interface: Touchscreen operation
- Customization: Appearance, panel marking and interface labeling can be discussed
IEC 60601-2-31 Defibrillation Pulse Tester Overview
The IEC 60601-2-31 Defibrillation Pulse Tester is a professional laboratory instrument used to evaluate the defibrillation protection performance of external cardiac pacemakers with internal power source. During clinical use, an external cardiac pacemaker may be exposed to defibrillation pulses. Therefore, the device must be tested to confirm that it can continue to meet basic safety and essential performance requirements after the specified pulse exposure.
The KP-900S uses a controlled RCL discharge circuit to generate defibrillation pulse waveforms and applies them to the equipment under test through a built-in three-chamber external pacemaker test network. This integrated design helps laboratories perform repeatable defibrillation protection testing according to IEC 60601-2-31 and YY 9706.231 requirements.
This product is a laboratory testing instrument. It is not a clinical defibrillator, not a cardiac pacemaker and not intended for patient treatment.
Applicable Standards and Test Method
The KP-900S is designed for IEC 60601-2-31:2020 and YY 9706.231-2023 related defibrillation protection testing of external cardiac pacemakers with internal power source. It supports standard pulse generation, polarity sequence testing and connection through the built-in three-chamber test network.
During testing, the external cardiac pacemaker is connected to the test network according to the laboratory procedure. The operator sets the required voltage, pulse polarity and pulse count through the touchscreen interface. The instrument then charges the internal storage capacitor and discharges through the RCL circuit to generate the required defibrillation pulse waveform.
The final sample connection, test clause, acceptance criteria and post-test functional evaluation should follow the applicable standard version, product risk analysis and laboratory quality procedure.
Technical Specifications
Detailed Pulse and Waveform Test Parameters
The following pulse output and waveform parameters are based on the KP-900S specification for IEC 60601-2-31 and YY 9706.231 defibrillation protection testing. Final test setup, polarity sequence, sample connection method and verification documentation can be reviewed according to the customer’s external cardiac pacemaker type, applicable standard version and laboratory procedure.
Standard and Test Scope
| Category | Specification | Notes |
|---|---|---|
| Product model | KP-900S | Defibrillation pulse tester |
| Applicable standards | IEC 60601-2-31:2020 / YY 9706.231-2023 | For external cardiac pacemaker defibrillation protection testing |
| Test object | External cardiac pacemakers with internal power source | EUT connection should follow laboratory procedure |
| Test purpose | Defibrillation protection testing | Verifies pulse withstand performance after exposure |
| Built-in test network | Three-chamber external pacemaker test network | Designed for the required test setup |
| Operation mode | Test mode / calibration mode | For routine testing and voltage verification |
Pulse Output and Waveform Parameters
| Parameter | Specification | Professional Notes |
| Pulse test circuit | RCL discharge circuit | Standard pulse generation circuit |
| Capacitance | 120 μF ±5% | Energy storage component |
| Inductance | 25 μH ±5% | Pulse waveform shaping component |
| Resistance condition | R_L + R = 1 Ω ±5% | Combined resistance condition of the discharge circuit |
| Maximum pulse voltage | 900 V | Tolerance: +2.5% / -0% |
| Energy range | 50 J to 55 J | Defibrillation pulse energy range |
| Waveform timing | Tw50: 4.05 ms to 4.6 ms | Time above 50% of peak voltage |
| Pulse polarity | Positive / negative polarity selectable | First pulse polarity can be set |
| Pulse sequence | 3 pulses of one polarity at 20–25 s intervals, ≥60 s pause, then 3 pulses of opposite polarity | Standard sequence for defibrillation protection testing |
| Test cycles | 1–999 times | Programmable test count |

Instrument Configuration
| Item | Specification | Notes |
| Control interface | Touchscreen operation | Parameter setting and status monitoring |
| Pulse output terminals | OutA / OutB | Pulse waveform output terminals |
| Test network interface | ME interface, Out1–Out5 | For three-chamber external pacemaker test network |
| Status indication | High-voltage output indicator | Helps identify output status during testing |
| Power supply | AC 220 V ±10%, 50/60 Hz standard configuration | Laboratory power configuration |
| Power consumption | <30 W | Low power benchtop design |
| Dimensions | 480 × 460 × 200 mm | Laboratory benchtop structure |
| Weight | Approx. 20 kg | Suitable for fixed laboratory use |
| Customization | Appearance, panel marking, interface labeling and logo position can be discussed | Must not affect standard-required waveform performance |
Technical Note : For IEC 60601-2-31 compliance testing, the RCL discharge circuit, pulse voltage, energy range, waveform timing, polarity sequence and built-in three-chamber test network should remain aligned with the applicable standard requirements. Appearance, panel marking, interface labeling, logo position or language display can be discussed according to customer laboratory needs, but such customization should not affect the standard-required pulse waveform, internal circuit parameters, safety protection design or verification performance.
Before formal testing, users should confirm the applicable standard version, test clause, sample connection method and waveform verification requirements. For laboratory quality control, regular verification of output voltage, pulse polarity, waveform timing and test network integrity is recommended.
Testing Principle
External cardiac pacemakers with internal power source may be exposed to defibrillation pulses during patient treatment. The IEC 60601-2-31 Defibrillation Pulse Tester uses an RCL discharge circuit to generate a controlled defibrillation pulse with defined voltage, energy and waveform timing.
The pulse is applied to the pacemaker under test through the built-in three-chamber external pacemaker test network. The test evaluates whether the pacemaker can maintain basic safety and essential performance after exposure to the required pulse sequence. By controlling pulse voltage, polarity, interval and test count, the KP-900S provides a repeatable method for defibrillation protection verification in medical device laboratories.
Built-in Test Network and Output Interface

The KP-900S integrates a built-in three-chamber external pacemaker test network for IEC 60601-2-31 related defibrillation protection testing. This helps reduce the need for additional external test network assembly and supports a more consistent laboratory setup.
The front panel includes the touchscreen interface, power switch, pulse output terminals, ME test network interfaces and high-voltage status indication. Operators can set test voltage, pulse polarity and pulse count, then monitor charging, discharging, pulse output and instrument status during the test process.
Key Features
The IEC 60601-2-31 Defibrillation Pulse Tester provides a dedicated RCL discharge circuit for external cardiac pacemaker defibrillation protection testing. It supports 900 V pulse output, 50 J to 55 J energy range, Tw50 waveform timing and selectable positive or negative pulse polarity.
The instrument integrates touchscreen operation, built-in three-chamber test network, pulse count setting, test mode, calibration mode and high-voltage output indication in one benchtop system. Its structure is suitable for medical device manufacturers, testing laboratories and quality inspection organizations that require repeatable IEC 60601-2-31 pulse testing.
Applications and Laboratory Use
The KP-900S is suitable for external cardiac pacemaker manufacturers, medical device testing laboratories, quality inspection institutions, R&D organizations and regulatory compliance teams.

Typical applications include defibrillation protection type testing, product development verification, sample comparison testing, production quality control, incoming inspection support and compliance preparation for external cardiac pacemakers with internal power source.
This product should not be used for defibrillation electrode performance testing. For disposable defibrillation electrodes or multifunction electrode performance evaluation, a different IEC 60601-2-4 related test system should be used.
Operation and Safety Notes
Before testing, users should confirm the applicable IEC 60601-2-31 or YY 9706.231 version, test clause, sample connection method, test voltage and pulse sequence requirement. The external cardiac pacemaker should be connected to the built-in test network according to the laboratory procedure before the pulse test starts.
Because the instrument generates high-voltage defibrillation pulses, it should only be operated by trained personnel in a controlled laboratory environment. Operators should confirm grounding, cable connection, output terminal condition, warning indicators and emergency safety procedures before testing.
The instrument is intended for laboratory testing only. It must not be connected to a patient or used for clinical defibrillation, pacing or treatment.
Calibration and Waveform Verification
- The KP-900S includes calibration mode for voltage verification and routine laboratory quality control. Users can verify output voltage, pulse polarity, pulse sequence and waveform timing according to the applicable calibration or verification procedure.
- Regular verification of the RCL discharge circuit, high-voltage output, waveform timing and built-in test network helps maintain repeatable defibrillation pulse test results. Calibration records, waveform verification records and inspection documentation can support laboratory quality management and compliance testing.
Download the KP-900S IEC 60601-2-31 Defibrillation Pulse Tester catalogue, including product overview, technical specifications and waveform reference.
Technical Inquiry & Expert Support
KingPo supplies the IEC 60601-2-31 Defibrillation Pulse Tester for defibrillation protection testing of external cardiac pacemakers with internal power source. Our engineering team can help review whether the KP-900S configuration matches your IEC 60601-2-31 test clause, sample connection method, waveform verification requirement and laboratory setup.
Please Provide the Following Information
- Applicable standard: IEC 60601-2-31:2020, YY 9706.231-2023 or internal test procedure
- Test object: External cardiac pacemaker with internal power source
- Related clause: Defibrillation protection test requirement, such as IEC 60601-2-31:2020
- Test requirement: RCL pulse test, polarity sequence, pulse count or waveform verification condition
- Connection requirement: Built-in three-chamber test network or special sample connection need
- Power supply: Required voltage and frequency for your laboratory
- Documentation: Calibration record, waveform verification record, inspection report, user manual or other quality document
- Customization needs: Appearance, front panel labeling, interface marking, logo position, language display, packaging or special laboratory layout requirement
Based on this information, KingPo can provide configuration suggestions, documentation support and quotation for your IEC 60601-2-31 Defibrillation Pulse Tester project.
Referenced Standards
* IEC 60601-2-31:2020 — Part 2-31: Particular requirements for the basic safety and essential performance of external cardiac pacemakers with internal power source
Defibrillation Pulse Tester Detail Display
IEC 60601-2-31 Defibrillation Pulse Tester FAQs
Related Products
Related News
DENG Sha¹, ZHANG Chao², WU Shaohai¹ Corresponding author: FAN Xiang¹ Guangdong Institute of Medical Device Quality Supervision and Inspection, No. […]
Conflict of Interest Disclosure: This is an independent, first-party engineering-style review based on Kingpo’s publicly available technical documentation (Tier 1) […]
Shan Chao 1 , Qiang Xiaolong 2 , Liu jiming 3 , Zhang Chao 3 . (1. Heilongjiang Institute for […]