




IEC 60601-2-31 Defibrillation Pulse Tester for External Cardiac Pacemakers
Designed for defibrillation protection testing of external cardiac pacemakers with an internal power source according to IEC 60601-2-31. It generates controlled defibrillation test pulses through an RCL discharge circuit to help verify whether the pacemaker can withstand defibrillation pulse stress and still maintain basic safety and essential performance after testing.
- Model: KP-900S
- Applicable standards: IEC 60601-2-31:2020 / YY 9706.231-2023
- Test object: External cardiac pacemakers with internal power source
- Test purpose: Defibrillation protection testing and pulse withstand verification
- Pulse circuit: RCL discharge circuit
- Maximum pulse voltage: 900 V
- Energy range: 50 J to 55 J
- Built-in test network: Three-chamber external pacemaker test network
- Control interface: Touchscreen operation
- Pulse sequence : 3 pulses of one polarity, ≥60 s pause, then 3 pulses of the opposite polarity
IEC 60601-2-31 Defibrillation Pulse Tester Overview
IEC 60601-2-31 Defibrillation Protection Test Pulse Generator
The KingPo KP-900S is a dedicated defibrillation protection test pulse generator for evaluating external cardiac pacemakers with an internal power source. It reproduces the specified defibrillation-related electrical stress condition so laboratories and manufacturers can verify whether the pacemaker remains safe and functional after pulse exposure.
The system generates the required pulse through an RCL discharge circuit comprising a 120 μF ±5% capacitor, 25 μH ±5% inductance and a combined resistance condition of RL + R = 1 Ω ±5%. It provides a maximum pulse voltage of 900 V, pulse energy from 50 J to 55 J and a Tw50 waveform timing range of 4.05 ms to 4.6 ms.
Unlike a general impulse generator or defibrillator analyzer, the KP-900S is configured specifically for IEC 60601-2-31 external cardiac pacemaker defibrillation protection testing. Its programmable pulse control, selectable first-pulse polarity and built-in pacemaker test network support controlled and repeatable laboratory verification.

Application and Output Summary
Technical Specifications
The following specifications define the standard KP-900S configuration for IEC 60601-2-31 and YY 9706.231 external cardiac pacemaker defibrillation protection testing.
| Model | KP-900S |
| Product Type | Defibrillation Protection Test Pulse Generator |
| Applicable Standards | IEC 60601-2-31:2020 / YY 9706.231-2023 |
| Test Object | External cardiac pacemakers with internal power source |
| Test Purpose | Defibrillation protection verification after pulse exposure |
| Operating Modes | Test mode and calibration mode |
| Pulse Test Circuit | RCL discharge circuit |
| Capacitance | 120 μF ±5% |
| Inductance | 25 μH ±5% |
| Resistance Condition | RL + R = 1 Ω ±5% |
| Maximum Pulse Voltage | 900 V, +2.5% / −0% |
| Pulse Energy | 50 J to 55 J |
| Tw50 Timing | 4.05 ms to 4.6 ms |
| First-Pulse Polarity | Positive or negative selectable |
| Pulse Sequence | 3 pulses of the selected polarity, followed by 3 pulses of the opposite polarity |
| Pulse Interval | 20 s to 25 s between pulses |
| Polarity Change Interval | At least 60 s before applying the opposite polarity |
| Built-in Test Network | Three-chamber external cardiac pacemaker test network |
| Related Test Arrangement | Supports the Figure 201.105-related test arrangement |
| Programmable Test Count | 1–999 cycles |
| Control Interface | Color touchscreen with programmable control |
| Pulse Output Terminals | OutA / OutB |
| Test Network Interfaces | ME interface and Out1–Out5 |
| Operating Functions | Voltage setting, charging, discharging, polarity selection, pulse counting and calibration |
| Protection Functions | Overvoltage protection and high-voltage output status indication |
| Parameter Memory | Test parameters retained after power interruption |
| Power Supply | AC 220 V ±10%, 50/60 Hz |
| Power Consumption | Less than 30 W |
| Dimensions | 480 × 460 × 200 mm |
| Weight | Approx. 20 kg |
Why Defibrillation Protection Testing Is Required
During clinical use, an external cardiac pacemaker may be exposed to defibrillation pulse stress through the patient connection path. The test determines whether the pacemaker can withstand the specified electrical stress without permanent damage.
After the required pulse sequence has been completed, the equipment is evaluated according to the applicable procedure to confirm that its basic safety and essential performance remain acceptable.
This is not simply a high-voltage output test. Pulse voltage, stored energy, RCL circuit parameters, waveform duration, polarity sequence, DUT connection and the pacemaker test network all affect the complete test condition.
RCL Defibrillation Pulse Generation
The KP-900S charges its energy-storage capacitor to the set voltage and then discharges it through the RCL circuit and pacemaker test network. The capacitance, inductance and resistance condition shape the pulse and establish the required voltage, energy and waveform timing.
Tw50 represents the interval during which the pulse voltage remains above 50% of its maximum value. The specified KP-900S Tw50 range is 4.05 ms to 4.6 ms.

Pulse Polarity and Timing Control
The operator can select the polarity of the first pulse group. The system applies three pulses of the selected polarity with an interval of 20 s to 25 s between each pulse.
After a pause of at least 60 s, the system applies three pulses of the opposite polarity. Programmable timing and pulse counting help reduce manual timing errors during repeated testing.

Built-in Pacemaker Test Network
The KP-900S includes a built-in three-chamber external cardiac pacemaker test network. The integrated network supports the required test arrangement and reduces the need to assemble a separate external network box for routine laboratory testing.
The instrument provides OutA and OutB pulse output terminals together with an ME interface and Out1–Out5 test network connections. DUT wiring should follow the instruction manual, applicable clause and laboratory procedure.
Test mode is used for normal pulse generation. Calibration mode allows output voltage checking, verification and compensation according to the laboratory’s calibration procedure.

Waveform, Voltage and Energy Verification
Available product records include measured pulse waveforms and an example measured energy value of approximately 52 J, providing a reference within the specified 50 J to 55 J output range.
For formal laboratory use, output voltage, pulse energy, Tw50 timing, polarity and test network integrity should be verified according to the laboratory’s quality system. Calibration documentation, waveform screenshots and inspection records can be confirmed according to project requirements.

Product Scope and Test Object
The KP-900S should be selected when the device under test is an external cardiac pacemaker with an internal power source and the required evaluation is defibrillation protection testing under IEC 60601-2-31 or YY 9706.231.
It is a laboratory test instrument for trained personnel. It is not a clinical defibrillator, cardiac pacemaker, patient-treatment device or general-purpose defibrillator analyzer, and it must not be connected to a patient.
The KP-900S is not intended for disposable defibrillation electrode pad testing. When the test object is a defibrillation electrode or multifunction electrode, use a dedicated IEC 60601-2-4 defibrillation electrode performance tester.
Typical Applications
Laboratory Safety and Configuration
The KP-900S generates high-voltage, high-energy pulses and should be operated only by trained laboratory personnel following the instruction manual, laboratory safety procedures and applicable standard requirements.
Before formal testing, users should confirm the applicable standard version, test clause, DUT connection method, pulse polarity sequence, waveform verification procedure and documentation requirements.
Appearance, panel markings, interface labels, display language and logo position may be discussed according to project requirements. Such customization must not change the RCL circuit parameters, pulse waveform, energy output, safety protection or verification performance.
Calibration and Waveform Verification
- The KP-900S includes calibration mode for voltage verification and routine laboratory quality control. Users can verify output voltage, pulse polarity, pulse sequence and waveform timing according to the applicable calibration or verification procedure.
- Regular verification of the RCL discharge circuit, high-voltage output, waveform timing and built-in test network helps maintain repeatable defibrillation pulse test results. Calibration records, waveform verification records and inspection documentation can support laboratory quality management and compliance testing.
Download Product Documentation
Download the KP-900S IEC 60601-2-31 Defibrillation Pulse Tester catalogue with technical specifications and waveform reference.
Technical Inquiry & Expert Support
KingPo supplies the IEC 60601-2-31 Defibrillation Pulse Tester for defibrillation protection testing of external cardiac pacemakers with internal power source. Our engineering team can help review whether the KP-900S configuration matches your IEC 60601-2-31 test clause, sample connection method, waveform verification requirement and laboratory setup.
Please Provide the Following Information
- Applicable standard: IEC 60601-2-31:2020, YY 9706.231-2023 or internal test procedure
- Test object: External cardiac pacemaker with internal power source
- Related clause: Defibrillation protection test requirement, such as IEC 60601-2-31:2020
- Test requirement: RCL pulse test, polarity sequence, pulse count or waveform verification condition
- Connection requirement: Built-in three-chamber test network or special sample connection need
- Power supply: Required voltage and frequency for your laboratory
- Documentation: Calibration record, waveform verification record, inspection report, user manual or other quality document
- Customization needs: Appearance, front panel labeling, interface marking, logo position, language display, packaging or special laboratory layout requirement
Based on this information, KingPo can provide configuration suggestions, documentation support and quotation for your IEC 60601-2-31 Defibrillation Pulse Tester project.
Referenced Standards
* IEC 60601-2-31:2020 — Part 2-31: Particular requirements for the basic safety and essential performance of external cardiac pacemakers with internal power source
Operation and Safety Notes
High-Voltage Pulse Safety
The KP-900S generates high-voltage pulse energy. Operators must avoid touching output terminals, test leads and connected circuits during charging, discharging or waiting-for-release status. The instrument should be operated only by trained personnel in a controlled laboratory environment.
Grounding Requirement
The instrument should be connected to a properly grounded power outlet. Proper protective grounding is required to reduce electric shock risk and support safe operation during high-voltage testing.
Test Connection Confirmation
Before each test, operators should confirm the EUT connection, output terminal polarity, ME test network connection, selected pulse polarity, test voltage and pulse count. Incorrect connection may damage the equipment under test or affect the validity of the test result.
Not for Clinical Use
The KP-900S must not be used for patient treatment, clinical defibrillation or cardiac pacing. It is only a laboratory pulse test instrument for medical electrical equipment testing.
Applications
External Cardiac Pacemaker Manufacturers
Manufacturers of external cardiac pacemakers can use the KP-900S for R&D verification, type testing preparation, internal compliance evaluation and product improvement. The tester helps evaluate whether the device can withstand defibrillation pulse stress without permanent damage.
Medical Device Testing Laboratories
Third-party medical device testing laboratories can use the KP-900S for IEC 60601-2-31 related test projects. The built-in test network, defined RCL circuit and programmable pulse output help support repeatable laboratory testing.
Certification and Compliance Preparation
Before submitting products for formal certification or regulatory testing, manufacturers can use the KP-900S for internal pre-compliance verification. This helps identify design risks related to defibrillation protection before the official test stage.
R&D and Quality Verification
R&D teams can use the tester to compare design changes, protection circuit performance and pacemaker response after defibrillation pulse exposure. Quality departments can use the instrument as part of internal verification for product safety testing.
Difference from Related Defibrillation Test Equipment
KP-900S Defibrillation Pulse Tester
The KP-900S is used for IEC 60601-2-31 external cardiac pacemaker defibrillation protection testing. It focuses on pulse generation and test network support for external cardiac pacemakers with internal power source.
Defibrillation Electrode Performance Tester
A defibrillation electrode performance tester is used for disposable defibrillation electrodes or multifunction electrode pads. It usually evaluates electrode impedance, defibrillation recovery, DC offset and pacing-related electrode performance. It is not the same test object as an external cardiac pacemaker.
ISO 14708-1 Defibrillation Test Pulse Generator
An ISO 14708-1 defibrillation test pulse generator is normally associated with active implantable medical device related testing. It should be selected according to the target standard and implantable device test requirement.
Defib-proof and Energy Reduction Tester
A defib-proof or energy reduction tester is used for other medical electrical equipment or applied part protection evaluations. It should not be confused with the KP-900S external cardiac pacemaker defibrillation pulse tester.
Related Products
| Related Product | Typical Application | Why It Is Relevant |
| IEC 60601-2-4 Defibrillation Electrode Performance Tester | Performance testing of disposable defibrillation electrodes and multifunction electrode pads. | Helps customers distinguish electrode performance testing from external cardiac pacemaker defibrillation protection testing. |
| Defibrillation Test Pulse Generator | Defibrillation pulse testing for active implantable medical device related applications. | Relevant for laboratories working with cardiac-related medical device pulse test requirements beyond IEC 60601-2-31. |
| Defib-proof and Energy Reduction Tester | Defibrillation protection and energy reduction testing for medical electrical equipment and applied parts. | Suitable for customers who need other defib-proof testing solutions in a medical electrical safety laboratory. |
| Dielectric Strength Tester for ISO 27186 | Dielectric strength and defibrillation waveform related testing of cardiac rhythm management connector systems. | Relevant for customers working with cardiac-related connector insulation and defibrillation waveform test requirements. |
Defibrillation Pulse Tester Detail Display
IEC 60601-2-31 Defibrillation Pulse Tester FAQs
Related Products
Related News
Medical Electrical Safety · Technical Delivery Case KP-DFT201 Defibrillation Effect and Energy Tester Delivered to DEKRA Guangzhou Laboratory KingPo delivered […]
Defibrillator Test Equipment Selection Guide A request for a “defibrillator tester” can describe several completely different laboratory tasks. The core […]
PROJECT CASE · AIMD DEFIBRILLATION IMMUNITY TESTING KP-1050S Defibrillation Test Pulse Generator Delivered to SGS Shanghai Laboratory KingPo delivered one […]
KingPo has supplied a KP-1050S Defibrillation Test Pulse Generator to NeuroPace, Inc. in Mountain View, California, United States. Project confirmed in […]
EU Regulatory Update · Medical Devices EU MDR Harmonised Standards 2026: Testing Updates for Medical Device Manufacturers EU MDR harmonised […]
Quick Answer Annual calibration of an ESU Analyzer is important because it helps ensure accurate electrosurgical unit testing, supports IEC […]
DENG Sha¹, ZHANG Chao², WU Shaohai¹ Corresponding author: FAN Xiang¹ Guangdong Institute of Medical Device Quality Supervision and Inspection, No. […]

