KingPo KP-1050S Delivered to SGS Shanghai Laboratory

Table of Contents

PROJECT CASE · AIMD DEFIBRILLATION IMMUNITY TESTING

KP-1050S Defibrillation Test Pulse Generator Delivered to SGS Shanghai Laboratory

KingPo delivered one KP-1050S Defibrillation Test Pulse Generator to SGS Shanghai Laboratory for controlled defibrillation pulse simulation in medical device safety testing.

The supplied system supports Test 1 damped sinusoidal defibrillation waveforms and Test 2 truncated exponential waveforms, with oscilloscope waveform observation, operation documents, matched accessories and calibration documentation prepared for the project.

KingPo KP-1050S system delivered to SGS Shanghai Laboratory for medical device safety testing
KP-1050S defibrillation pulse generator prepared for medical device safety testing.

Project Overview

Customer
SGS Shanghai Laboratory
Equipment
KP-1050S
Application
Defibrillation Pulse Simulation
Documentation
Calibration Report
Supplier KingPo Technology Development Limited
Delivered Equipment KP-1050S Defibrillation Test Pulse Generator
Main Test Purpose Controlled Test 1 and Test 2 defibrillation pulse simulation for AIMD-related medical device safety testing
Related Standards GB 16174.1-2024 20.2, ISO 14117:2012, ISO 14708-1:2003, EN 45502-1:2015, EN 45502-2-2:2008
Project Support Matched accessories, operation documents, setup support, oscilloscope waveform observation and calibration documentation

Project Background

Active implantable medical devices, such as pacemakers and implantable cardioverter defibrillators, may be exposed to external defibrillation pulses during clinical use. For this reason, laboratories need controlled defibrillation pulse simulation equipment to evaluate whether the device under test can maintain its defined safety and functional condition after specified pulse exposure.

The KP-1050S was supplied for a laboratory workflow where waveform type, output voltage level, polarity, test sequence, DUT connection and waveform observation all need to be clearly defined. Compared with a simple high-voltage pulse source, this type of medical device test system must support repeatable pulse generation, documented operation and safe setup control.

Customer Testing Requirement

For defibrillation immunity testing, the pulse generator is only one part of the complete setup. Before routine use, laboratories normally need to confirm the applicable standard, DUT category, waveform requirement, output polarity, load condition, connection method, pulse duration, waveform observation method, operator protection and documentation requirements.

Test 1 Requirement

Damped sinusoidal defibrillation waveform with 140 V / 380 V voltage levels.

Test 2 Requirement

Truncated exponential defibrillation waveform with 140 V / 270 V voltage levels and single-phase or dual-phase selection.

Waveform Observation

Use a digital oscilloscope or suitable measuring setup to observe output waveform behavior during setup.

Laboratory Setup and Waveform Observation

During the setup support process, the KP-1050S was connected with a digital oscilloscope for waveform observation. The setup images show the practical connection between the pulse generator and measuring equipment, helping technical buyers understand how output behavior can be reviewed before routine testing.

KP-1050S and oscilloscope setup for medical device safety testing
Laboratory setup with oscilloscope for defibrillation pulse testing.
KingPo KP-1050S connected with oscilloscope for waveform verification
KP-1050S connected with oscilloscope for waveform verification.
Defibrillation pulse waveform measurement gallery with oscilloscope close-up images
Oscilloscope waveform measurement images from the test setup.

KP-1050S Product Capability Used in This Project

The KP-1050S integrates two defibrillation pulse test modes into one instrument. It is designed for AIMD-related defibrillation immunity testing where laboratories need to switch between damped sinusoidal and truncated exponential waveform applications, manage polarity, observe capacitor voltage and record pulse operation in a controlled workflow.

Capability Area KP-1050S Configuration
Test 1 Waveform Damped sinusoidal defibrillation waveform, 140 V / 380 V
Test 2 Waveform Truncated exponential defibrillation waveform, 140 V / 270 V, single-phase or dual-phase selectable
Polarity Control Positive and negative polarity can be selected according to the test sequence
Automated Sequence Typical sequence includes three positive pulses, a 60 s pause and one negative pulse
Programmable Cycles 1–999 programmable test counts
Operation Interface Touchscreen interface for voltage setting, charge/discharge control, polarity selection, pulse count, timer and status monitoring
External Network KP-1050-1 resistor network box for the related test setup
Calibration Support Calibration interface with positive and negative voltage compensation functions

Technical Specification Reference

Applicable Standards GB 16174.1-2024 20.2, ISO 14117:2012, ISO 14708-1:2003, EN 45502-1:2015, EN 45502-2-2:2008
Test 1 Voltage 140 V / 380 V, tolerance +5% / -0%
Test 1 Waveform Timing 1.5 ms < Tp < 2.5 ms; 3 ms < Tw50 < 5.5 ms
Test 2 Voltage 140 V / 270 V, tolerance +5% / -0%
Test 2 Waveform Timing 9.5 ms < Td < 10.5 ms; 1 μs < tr < 5 μs; 1 μs < tf < 5 μs
Test Count 1–999 times
Input Power AC 220 / 110 V ±10%, 50 / 60 Hz, max power approx. 300 W
Dimensions 480 × 460 × 200 mm
Weight Approx. 10 kg

Final configuration should be confirmed according to the approved technical specification, calibration record and laboratory test plan for the supplied system.

Test Setup Workflow for Similar Projects

A defibrillation pulse generator project usually requires more than equipment delivery. The laboratory needs to confirm the test objective, connect the DUT safely, verify the waveform and prepare operating records before routine testing.

  1. Confirm applicable standard, DUT category and required pulse mode.
  2. Prepare connection accessories, load condition and safety spacing.
  3. Connect the KP-1050S with the measurement setup for waveform observation.
  4. Check output waveform behavior, polarity and pulse sequence.
  5. Record setup photos, operation records and calibration documentation.

KP-1050S Test 2 waveform shown on oscilloscope during laboratory setup
KP-1050S Test 2 waveform and generator setup during laboratory verification.

Calibration Documentation

The project includes calibration documentation to support laboratory equipment records, internal review, traceability management and medical device safety testing workflows. The KP-1050S also provides calibration interface functions for voltage compensation during equipment management.

View KP-1050S Calibration Report

Applicable Standards and Test Context

The KP-1050S is designed for defibrillation pulse testing applications related to active implantable medical device safety evaluation. The related test method depends on the DUT category, standard edition, laboratory procedure and project scope.

Standard Relationship to the KP-1050S Application
GB 16174.1-2024 20.2 Application context for active implantable medical device defibrillation-related testing.
ISO 14117:2012 Defibrillation immunity testing for active implantable medical devices.
ISO 14708-1:2003 General safety requirements context for active implantable medical devices.
EN 45502-1:2015 Safety requirements context for active implantable medical devices.
EN 45502-2-2:2008 Application context related to implantable cardioverter defibrillator testing.

Engineering Notes for Similar Test Setups

For defibrillation-related testing, laboratories should not select equipment only by product name. The complete setup should be reviewed according to the DUT type, standard, intended test method, waveform requirement, connection method and documentation requirements.

Before Quotation

Confirm applicable standard, DUT type, Test 1 or Test 2 requirement, voltage level, accessories and calibration requirement.

Before Routine Testing

Verify waveform behavior, connection method, load condition, polarity, safety spacing and operation sequence.

For Documentation

Prepare operation records, setup photos, waveform records, calibration documents and internal laboratory procedures.

Project Value

The delivered KP-1050S setup is intended to support controlled, repeatable and documented defibrillation pulse simulation in laboratory testing workflows. The integration of Test 1 and Test 2 waveforms, touchscreen operation, polarity control, oscilloscope waveform observation and calibration documentation helps engineers prepare a clearer and more traceable test process.

  • One system for Test 1 and Test 2 defibrillation pulse applications
  • Touchscreen operation for voltage setting and charge/discharge control
  • Oscilloscope waveform observation during setup support
  • Calibration documentation for equipment records

Related Product and Technical Pages

KP-1050S Defibrillation Test Pulse Generator
Main product page for the delivered equipment.

ISO 14708-1 Clause 20.1 Test 1/2 Defibrillation Test Pulse Generator
Detailed KP-1050S application and specification page for AIMD testing.

Medical Test Equipment
Broader KingPo medical device safety testing equipment category.

ISO 14117 Defibrillation Test Pulse Generator Guide
Technical article explaining pulse testing for active implantable medical devices.

FAQ

What did KingPo deliver to SGS Shanghai Laboratory?

KingPo delivered one KP-1050S Defibrillation Test Pulse Generator with matched accessories, operation documents, setup support, oscilloscope waveform observation and calibration documentation.

What waveforms can the KP-1050S generate?

The KP-1050S supports Test 1 damped sinusoidal defibrillation waveforms and Test 2 truncated exponential defibrillation waveforms. Test 2 can be configured as single-phase or dual-phase according to the test requirement.

Which standards are related to this testing?

The related standards include GB 16174.1-2024 20.2, ISO 14117:2012, ISO 14708-1:2003, EN 45502-1:2015 and EN 45502-2-2:2008. The final test method depends on the DUT category and laboratory test plan.

Why is oscilloscope waveform observation important?

Oscilloscope waveform observation helps laboratory engineers review generated pulse behavior, check polarity and waveform shape, and record setup conditions before routine testing.

What should customers confirm before selecting a similar system?

Customers should confirm the applicable standard, DUT type, Test 1 or Test 2 requirement, voltage level, load condition, pulse polarity, connection method, accessories, calibration requirement and safety operation procedure.

Is calibration documentation available for KP-1050S projects?

Yes. Calibration documentation can support equipment records, traceability management, internal review and laboratory quality procedures. The KP-1050S also includes calibration interface functions for voltage compensation.

Need a similar defibrillation pulse testing setup?

KingPo can help confirm the applicable standard, DUT type, Test 1 / Test 2 waveform requirement, accessories, documentation, calibration report needs and operation guidance requirements before quotation.

Contact KingPo

Picture of Bruce Zhang

Bruce Zhang

Bruce Zhang is the Founder and Senior Engineer of KingPo Technology Development Limited, with over 16 years of experience in environmental and safety testing technologies. As a member of SAC TC118, TC338, and TC526, he participates in national standard reviews and provides technical guidance on IEC and ISO compliance for global laboratories.

Welcome To Share This Page:

Related Products

Related News

PROJECT CASE · AIMD DEFIBRILLATION IMMUNITY TESTING KP-1050S Defibrillation Test Pulse Generator Delivered to SGS Shanghai Laboratory KingPo delivered one

KingPo has supplied a KP-1050S Defibrillation Test Pulse Generator to NeuroPace, Inc. in Mountain View, California, United States. Project confirmed in

Ensure ISO 14117 compliance for AIMDs by using a Defibrillation Test Pulse Generator to meet safety standards and document reliable defibrillation immunity testing.
Ensure IEC 60601 compliance in 2026 with Defib-proof and Energy Reduction Tester for safer, faster medical device approval and reliable patient protection.

In the realm of electrical safety, where materials must endure potential fire risks, the Kingpo KP-FT01 Glow Wire Tester emerges

Glow Wire Tester The Kingpo KP-FT01 glow wire tester is a high-precision, IEC 60695-2-10 compliant apparatus designed for reliable fire

The glow wire test (GWT) is a cornerstone fire hazard assessment method in electrical and electronic product safety, as defined

The Defibrillation Immunity Testing Requirements for Active Implantable Medical Devices (AIMDs): Standards Breakdown, Visual Insights, and Recommendation for the KP-1050S

Scroll to Top

Get A Free Quote Now !

Contact Form
If you have any questions, please do not hesitate to contact us.