KP-1050S Defibrillation Test Pulse Generator for Medical Device Safety
KP-1050S Defibrillation Test Pulse Generator for Medical Device Safety

KP-1050S Defibrillation Test Pulse Generator for Medical Device Safety

The KP-1050S defibrillation test simulator ensures medical device safety compliance, offering reliability and precise electrical performance verification.

  • Standard Compliance: IEC 60601-1 – Direct reference for defibrillation pulse testing
  • Pulse Simulation: Accurate defibrillation waveform generation
  • High Voltage Capability: Safe, repeatable high-energy pulse delivery
  • Durable Construction: Robust design for medical lab reliability

MOQ: 1

Price: To be quoted

Payment Method: T/T

Product Description

The KP-1050S defibrillation test pulse generator is used to evaluate whether active implantable medical devices will be permanently affected when subjected to damped sine defibrillation waveforms and truncated exponential defibrillation waveforms , whether they can be reprogrammed to restore settings , and to evaluate the safety and programmability of the product. The KP-1050S defibrillation test pulse generator complies with ISO14708-1 and GB 16174.1-2024 20.2 standards. The following test procedures are included:

  • Test 1

1‌. Test Purpose‌ :

To evaluate whether active implantable medical devices are permanently affected when subjected to damped sinusoidal defibrillation waveforms and whether they can be reprogrammed to restore settings.

2. Test equipment :

Defibrillation test voltage generator, used to provide a damped sinusoidal defibrillation waveform with specific characteristics.

3. Test parameters :

The parameters of the damped sinusoidal defibrillation waveform must meet the following requirements: 1.5ms < Tp < 2.5ms; 3ms < Tw50 < 5.5ms; where Tp is the time interval from the start of the waveform to the maximum value, and Tw50 is the time interval above 50% of the maximum value.

The maximum value of the test voltage Vtest is adjusted to 140+70V.

  • Test 2

1. Purpose of the study :

To further evaluate the safety and reprogrammability of active implantable medical devices when subjected to truncated exponential defibrillation waveforms.

2. Test equipment :

A defibrillation pulse generator is used to provide a truncated exponential defibrillation waveform of a specific duration.

‌3. Test arrangement‌ :

Test arrangement using C1=(150±50)μF, R1=65Ω and 2 sets of coupled switches S1 and S2.

A single-phase or bi-phase truncated exponential defibrillation waveform with a duration of Td = (10 ± 0.5) ms is generated between output terminals A and B.

4. Test parameters :

The test voltage-time parameters of the truncated exponential defibrillation waveform must meet the numerical range in Table 1.

The maximum value of the test voltage Vtest is adjusted to 140+70V.

This model of pulse tester uses a programmable controller as its core and is equipped with a color touch screen for intuitively displaying and setting all working parameters and instrument status. Its user-friendly design makes the operation interface concise and clear, and it can memorize various parameters even in the case of power failure, greatly improving the convenience of use. In terms of safety performance, this model of pulse tester has a built-in overvoltage protection device, which effectively guarantees the stable operation of the instrument during the test and ensures the safety and reliability of the test work. In addition, it also has the characteristics of easy operation, excellent performance, high reliability and good cost performance. Therefore, it is deeply favored by electrical safety testing laboratories and electronic and electrical appliance manufacturers, and has become their preferred tool for electrical safety testing.

Detail Information

Place of Origin China Brand Name Kingpo
Certification iso9001 ce Model Number KP-1050S
Voltage Measurement Accuracy: AC 110/220V Input Power: AC 110/220V ±10%, 50/60Hz, 425W
Weight: 22.5 Kg
Highlight: AC 110/220V Test Pulse Generator, 425W Test Pulse Generator, Implantable Medical Device Test Pulse Generator

Get Free Quote of KP-1050S Defibrillation Test Pulse Generator for Medical Device Safety

We provide high-precision Defibrillation Test Pulse Generator direct from KingPo factory, ensuring reliable IEC 60601-1 medical safety testing.

The Core Value We Deliver to You

  • Exact IEC 60601-1 compliance – Accurate defibrillation pulse simulation
  • High Reliability – Robust design for repeatable results
  • Factory Advantages – MOQ 1 unit, 15-day delivery, 1-year warranty, ISO9001 & CE
  • Complete Testing Support – Pair with our IEC series for full checks
  • Faster Certification – Identify safety risks early, reduce failures

Take Action Now – Submit inquiry in 30 seconds for best price, stock status, and technical support within 24 hours!

Technical Indicators

Technical Projects Parameter
Comply with standards ISO 14708-1:2014 & GB 16174.1-2024 20.2
Test 1 voltage 140+70V
Test 1 Circuit RLC Circuit
Damped sine defibrillation waveform 1.5ms < Tp < 2.5ms; 3ms < Tw50 < 5.5ms
Output polarity Positive and negative polarity output can be freely set
Test 1 Program First send 3 positive voltage pulse sequences (each pulse interval is 20+2s), then send a negative voltage pulse after an interval of 60+2s
Test 2 voltage 140+70V
Phase switching Single phase/double phase can be set freely
Monophasic/Biphasic Truncated Exponential Defibrillation Waveform 9. 5ms<Td<10.5ms
1μs<tr<5μs
1μs<tf<5μs
Test 2 Program First send 3 positive polarity single-phase voltage pulse sequences (each pulse interval is 20+2s), then send a negative polarity voltage pulse after an interval of 60+2s
‌Evaluation Criteria‌ Active implantable medical devices are tested for compliance by checking that they have not been permanently affected and can be reprogrammed to restore settings after completing the steps of Test 1 and Test 2.
Number of trials 1 – 999 times
Pulse generator output mode Floating output
Input Power AC 220V±10%, 50Hz/60Hz
Dimensions 680*460*200mm
weight About 30 kg

Test Waveform

FAQ

How does the Defibrillation Test Pulse Generator simulate real defibrillation pulses?
It generates accurate waveforms per IEC 60601-1 to test insulation withstand and leakage current, ensuring medical devices survive emergency defibrillation.
Can this tester be used for both defibrillators and patient monitors?
Yes – it fully complies with IEC 60601-1 requirements for patient-connected devices, covering insulation and leakage testing.
What if my lab needs high-voltage safety – is the tester secure?
Yes – it features protected circuits and safe operation design, ideal for medical safety labs.
How does precise pulse simulation help reduce certification failures?
Accurate defibrillation pulse delivery minimizes measurement errors, providing reliable data for IEC 60601-1 compliance and reducing re-test risks.
What is the MOQ and delivery time for the Defibrillation Test Pulse Generator?
MOQ 1 unit; delivery typically 15 days, allowing quick setup for urgent testing needs.
What warranty and support does KingPo offer?
1-year warranty plus lifetime technical support for long-term reliability.
How does this tester help manufacturers improve patient safety?
It identifies insulation weaknesses early, preventing leakage or breakdown during real defibrillation, reducing risks.
Can the tester be used for custom pulse testing in R&D?
Yes – adjustable parameters allow customization for different medical device designs.
How long does a typical defibrillation pulse test cycle take?
A single cycle is quick (minutes), saving lab time compared to less specialized equipment.
How to get a quote or sample for the Defibrillation Test Pulse Generator?
Submit the inquiry form or email [email protected] for pricing and fast support.

Related Products

Related News

In the complex world of electrical engineering and manufacturing, ensuring the safety of electrical components is paramount. One of the […]

Ensure ISO 14117 compliance for AIMDs by using a Defibrillation Test Pulse Generator to meet safety standards and document reliable defibrillation immunity testing.
Ensure IEC 60601 compliance in 2026 with Defib-proof and Energy Reduction Tester for safer, faster medical device approval and reliable patient protection.

For plastic insulation frames, given the high heat generated by the copper wires and iron core, if the flame retardant […]

The glow wire test (GWT) is a cornerstone fire hazard assessment method in electrical and electronic product safety, as defined […]

IP ratings explained: Understand each code, what the digits mean, and how to choose the right protection level for electronics in any environment.
Scroll to Top

Get A Free Quote Now !

Contact Form
If you have any questions, please do not hesitate to contact us.