Liquid particle counter for evaluating particulate matter released from cochlear implant components under ISO 14708-7:2019 Clause 14.2.
Liquid particle counter for evaluating particulate matter released from cochlear implant components under ISO 14708-7:2019 Clause 14.2.
Liquid particle testing equipment for controlled preparation of reference and test samples used in particulate matter evaluation under ISO 14708-7:2019.
Optical liquid particle analyzer for measuring particles in liquid samples used in ISO 14708-7:2019 related particulate matter testing.
Laboratory liquid particle analyzer for particulate matter testing of implantable medical device components according to ISO 14708-7:2019 Clause 14.2.

ISO 14708-7 Liquid Particle Counter for Cochlear Implant Particulate Matter Testing

The Liquid Particle Counter is designed to support particulate matter testing associated with ISO 14708-7:2019, Clause 14.2, which addresses unintended particulate matter released from implantable parts of cochlear and auditory brainstem implant systems. The system integrates optical particle detection, quantitative liquid sampling, synchronized reference and test sample stirring, temperature control and test data management.

With a specified particle detection range of 1–500 μm, the system supports particle counting at ≥5 μm, ≥10 μm and ≥25 μm thresholds. It helps laboratories compare particle counts in test and reference liquids under controlled sample preparation conditions.

  • Product Type: Liquid Particle Counter / Particulate Matter Testing System
  • Model: KP-MP500D
  • Primary International Standard: ISO 14708-7:2019, Clause 14.2
  • Test Application: Particulate matter release testing for cochlear implant systems
  • Particle Detection Range: 1–500 μm
  • Particle Counting Thresholds: ≥5 μm, ≥10 μm and ≥25 μm
  • Minimum Selectable Sampling Volume: 0.1 mL
  • Sample Processing: Synchronized reference and test sample stirring
  • Temperature Control: 37°C ± 2°C
  • Typical Users: Medical device manufacturers, R&D laboratories, quality control departments and third-party testing laboratories

ISO 14708-7 Liquid Particle Counter

Product Overview

Particulate matter released from implantable medical device components is an important consideration in evaluating device safety. ISO 14708-7:2019 includes requirements for assessing unintended particulate matter released from implantable parts intended to contact body fluids during normal use.

The KP-MP500D Liquid Particle Counter combines controlled sample processing with quantitative particle analysis to support this evaluation. Its dual-sample arrangement enables simultaneous stirring of reference and test samples, while temperature control and adjustable stirring speed support consistent sample preparation.

The system provides particle counts at configured size thresholds and supports selectable sampling volumes. Its control software includes test method configuration, calibration management, historical result queries and data output for laboratory testing workflows.

Liquid particle counter in a laboratory for medical device particulate matter testing related to ISO 14708-7:2019 Clause 14.2.

Technical Specifications

Parameter Specification
Product Name Liquid Particle Counter for Particulate Matter Testing
Model KP-MP500D
Primary International Standard ISO 14708-7:2019, Clause 14.2
Related Chinese Standard YY 0989.7—2017
Particle Detection Range 1–500 μm
Particle Counting Thresholds ≥5 μm, ≥10 μm and ≥25 μm
Particle Counting Range 0–9,999,999 particles
Particle Concentration Range 0–10,000 particles/mL
Relative Standard Deviation (RSD) ≤2% at a standard particle concentration of ≥1,000 particles/mL
Accuracy ±5%
Channel Resolution ≥95% for the ≥10 μm channel
Minimum Selectable Sampling Volume 0.1 mL
Sampling Time — 5 mL Less than 10 s
Sampling Time — 100 mL Less than 3 min
Stirring Speed 0–1500 r/min
Temperature Control 37°C ± 2°C
Sample Processing Simultaneous stirring of reference and test samples
Custom Test Settings At least 2 sets
Detection System Optical particle detection
Additional Test Function 20 μm particle removal efficiency testing
Software Functions Test method configuration, calibration management, maintenance, historical queries and result management

ISO 14708-7:2019 Particulate Matter Testing

ISO 14708-7:2019 :
Implants for Surgery — Active Implantable Medical Devices — Part 7: Particular Requirements for Cochlear and Auditory Brainstem Implant Systems,
specifies particular requirements for these active implantable medical devices.

Clause 14.2 addresses the release of unintended particulate matter from implantable parts. The associated test method evaluates the particle count in a test liquid relative to a reference liquid after exposure under specified conditions.

Test Conditions and Acceptance Criteria

The standard’s particulate matter test procedure includes saline immersion, controlled temperature and continuous agitation.

Test Item Requirement
Test Medium Saline solution, approximately 9 g/L
Temperature 37°C ± 2°C
Agitation Period 8–18 h
Evaluation Method Compare the test sample with a reference sample
Particles Greater Than 5.0 μm Excess average count not exceeding 100 particles/mL
Particles Greater Than 25.0 μm Excess average count not exceeding 5 particles/mL
  • Liquid particle counter interface displaying total counts and particle-size channels for particulate matter analysis associated with ISO 14708-7:2019.

The specified limits apply to the excess average particle count relative to the reference sample, rather than the absolute particle concentration measured by the instrument.

The KP-MP500D supports the relevant sample preparation and particle counting operations. For test procedures requiring extended continuous agitation, the required operating duration should be included in the equipment configuration requirements.

Test Principle and Workflow

The KP-MP500D combines controlled sample preparation with optical particle detection to support comparative particulate matter analysis.

01Sample Preparation

Prepare the test specimen and reference sample according to the applicable test procedure, including the specified liquid and specimen preparation conditions.

02Parameter Configuration

Set the required sampling volume, particle counting thresholds, stirring speed and temperature.

03Synchronized Sample Processing

Stir the reference and test samples simultaneously under the specified preparation conditions.

04Quantitative Sampling

Collect the required liquid sample volume for particle measurement.

05Particle Counting

Use the optical detection system to obtain particle counts at the configured size thresholds.

06Comparative Evaluation

Compare the test and reference results using the calculation method and acceptance criteria specified by ISO 14708-7:2019.

07Result Management

Store and retrieve test methods, calibration information and measurement results through the control software.

Key Features

Synchronized Reference and Test Sample Processing

Simultaneous stirring helps maintain comparable preparation conditions for the reference and test liquids, supporting consistent comparative particle analysis.

Particle Counting Across Multiple Size Thresholds

The specified 1–500 μm detection range and three counting thresholds support size-based particle analysis for relevant medical device testing procedures.

Quantitative Liquid Sampling

Selectable sampling volumes starting at 0.1 mL accommodate different sample volume requirements. Specified sampling times are less than 10 seconds for 5 mL and less than 3 minutes for 100 mL.

 

  • Liquid particle counter test method selection interface for configuring laboratory particle testing procedures, including custom test settings.

Temperature-Controlled Stirring

Temperature control at 37°C ± 2°C and an adjustable stirring speed of 0–1500 r/min support controlled sample preparation for relevant particulate matter test methods.

Calibration and Test Data Management

The software supports test method configuration, calibration management, historical result queries and data output, helping laboratories organize measurement records and repeat configured test methods.

Typical Applications

Application Testing Purpose
Cochlear Implant Development Evaluate particulate matter released from relevant implantable components
Medical Device Verification Compare particle counts between test and reference samples
Quality Control Laboratories Review particle counts and size-based measurement results
Third-Party Testing Laboratories Support particulate matter testing under applicable medical device standards
R&D Laboratories Investigate particulate release under defined sample preparation conditions

The primary application is particulate matter testing associated with cochlear implant systems. Other implantable medical device applications should be assessed against the relevant standard and test procedure.

Compliance & Regulatory Assurance

The Liquid Particle Counter is designed to support particulate matter testing in accordance with ISO 14708-7:2019, Clause 14.2. YY 0989.7—2017 is the corresponding Chinese industry standard reference.

The applicable test conditions and equipment configuration should be confirmed against the required standard and testing procedure.

Configuration and Selection Guide

To determine the appropriate test setup, please provide the following information:
Information Details to Provide
Standard & Test Method Applicable standard and clause, such as ISO 14708-7:2019 Clause 14.2, including any specific laboratory test procedure.
Specimen & Preparation Device or component type, dimensions, exposed surface area and sample preparation requirements.
Particle Measurement Required particle-size thresholds and sampling volume, including any requirements beyond the specified measurement range.
Additional Testing Specify whether 20 μm particle removal efficiency testing or particular reporting requirements are needed.

These details help determine a suitable configuration for the intended particulate matter testing procedure.

ISO 14708-7 Liquid Particle Counter

FAQ

What does the liquid particle counter measure under ISO 14708-7:2019 Clause 14.2?
It measures particle counts in the saline extract from an implantable component for comparison with a reference sample. The results help evaluate unintended particulate matter released during the specified test procedure.
What saline volume is required for ISO 14708-7 Clause 14.2 testing?
The required saline volume is 5 ± 0.5 mL per cm² of the implantable part's surface area. The specimen and reference sample must be prepared according to the standard's specified conditions.
What temperature and agitation duration does ISO 14708-7 specify?
The test bath must be maintained at 37°C ± 2°C for 8–18 hours, with continuous agitation throughout the exposure period. The reference sample must be maintained and agitated under similar conditions.
Why is a reference sample required for particulate matter testing?
The reference sample establishes the background particle count of the test liquid. The specimen liquid is compared with a reference prepared from the same batch of saline to determine the excess average particle count.
What are the particle count acceptance criteria in ISO 14708-7 Clause 14.2?
The excess average particle count in the specimen sample must not exceed 100 particles/mL for particles greater than 5.0 μm or 5 particles/mL for particles greater than 25.0 μm. These limits apply to the comparison with the reference sample.
Which particle sizes can the KP-MP500D measure?
The KP-MP500D has a specified particle detection range of 1–500 μm, with counting thresholds of ≥5 μm, ≥10 μm and ≥25 μm. The selected measurement settings and reporting method should match the particle-size definitions in the applicable test procedure.
Does the 0.1 mL minimum sampling volume determine the saline immersion volume?
No. The minimum selectable sampling volume is an instrument parameter. The saline immersion volume for ISO 14708-7 Clause 14.2 is calculated from the implantable part's surface area, independently of the volume subsequently sampled for particle counting.
Is 20 μm particle removal efficiency a requirement of ISO 14708-7 Clause 14.2?
No. Clause 14.2 specifies comparative particle count acceptance criteria for particles greater than 5.0 μm and 25.0 μm. The 20 μm particle removal efficiency function is a separate measurement application.
What should be confirmed before selecting a particle counter for ISO 14708-7 testing?
Confirm the required particle-size measurement method, counting thresholds, sample volume, reference sample arrangement and continuous agitation duration. The equipment configuration should support the complete test procedure, including the specified 8–18-hour exposure period.

Related Products

Related News

The International Electrotechnical Commission (IEC) published Amendment 2 to IEC 60601-2-4 on September 28, 2026. The amendment concerns the particular […]

IEC 60601-2-2, which specifies particular requirements for the basic safety and essential performance of high-frequency surgical equipment and accessories, is […]

Defibrillation electrode testing verifies whether disposable and multifunction electrode pads can maintain stable impedance, recover after a 360J shock, control […]

Choose a medical bed side rail strength tester for IEC 60601-2-52 by focusing on force range, accuracy, compliance, and safety features.
Comparing IEC 60601-2-34 and IEC 60601-2-25: A Technical Guide details device scope, testing, and compliance differences for medical device standards.
KINGPO’s Electrosurgical Unit analyzer ensures ESU safety and IEC 60601-2-2 compliance with precise testing, supporting patient protection and regulatory needs.
Ensure IEC 60601 compliance in 2026 with Defib-proof and Energy Reduction Tester for safer, faster medical device approval and reliable patient protection.

Technical Insight for IEC 60601-2-2 High-Frequency Surgical Equipment Testing High-frequency electrosurgical units (ESU) operate in the radio frequency range above […]