Active Implantable Medical Device Pressure Testing System

KingPo KP3105 is a programmable pressure testing system designed to evaluate the resistance of active implantable medical devices to atmospheric pressure changes. It supports pressure cycling, controlled high-pressure exposure and absolute-pressure holding tests for devices such as cochlear implants, auditory brainstem implants and cardiac pacemakers. The system uses a water pressure chamber to apply controlled pressure profiles and record pressure-time data, supporting test procedures associated with ISO 14708-7, ISO 14708-1 and ISO 14708-2.

  • Model: KP3105
  • Product Type: Active Implantable Medical Device Pressure Testing System
  • Primary Use: Pressure resistance, pressure cycling and absolute-pressure holding tests.
  • Applicable Standards: ISO 14708-7:2019, ISO 14708-1:2014, ISO 14708-2:2019, GB 16174.1-2024 and GB 16174.2-2024.
  • Key Test Capability: High-pressure capability ≥304 kPa; pressure cycling ≥40 cycles; pressure change rate ≥152 kPa/min.
  • Pressure Holding: 150 kPa ±7.5 kPa absolute pressure for at least 1 hour.
  • Test Chamber: Transparent water pressure chamber with an effective volume of ≥500 mL.
  • Data Management: Programmable test sequences, pressure-time recording and USB/LAN data export.
  • Typical DUTs: Cochlear implant systems, auditory brainstem implant systems and cardiac pacemakers.

Active Implantable Medical Device Pressure Testing System

Product Overview

Active implantable medical devices may be exposed to different atmospheric pressure conditions during transportation and normal use. Pressure variations can affect the structural integrity and functional stability of implantable components.

The KingPo KP3105 provides a controlled pressure environment for evaluating these effects through programmed pressure changes, repeated pressure cycles and sustained pressure exposure.

The system is designed for laboratories performing pressure-related verification of active implantable medical devices, with configurable test profiles based on applicable standard requirements and DUT specifications.

Technical Specifications

The following specifications summarize the pressure-control capability, chamber configuration and data management functions of the KP3105.

Parameter Specification
Model KP3105
Equipment Type Active Implantable Medical Device Pressure Testing System
Test Medium Water
High-Pressure Capability ≥304 kPa
Pressure Change Rate ≥152 kPa/min
Pressure Cycling ≥40 cycles
Pressure Ramp-Up Time ≤2 min
Pressure Ramp-Down Time ≤2 min
High-Pressure Holding Time ≥2 min
Absolute-Pressure Holding 150 kPa ±7.5 kPa
Absolute-Pressure Holding Duration ≥1 hour
Chamber Effective Volume ≥500 mL
Chamber Structure Transparent water pressure chamber
Chamber Fastening Multi-point fastening structure
Specimen Fixture Stainless steel fixtures and positioning supports
Control Method Programmable automatic pressure control
Status Display Current pressure, operating stage and cycle number
Data Recording Pressure-time data
Communication Interfaces USB, LAN

Engineering Note: The pressure reference used for the equipment’s rated pressure and control display must be confirmed during final configuration. The applicable ISO test requirements specify absolute-pressure values.

Internal electrical control components of the KingPo KP3105 implantable medical device pressure testing system

Applicable Standards and Test Requirements

The KP3105 is designed to support pressure-related testing specified in the ISO 14708 series, covering different categories of active implantable medical devices. The applicable test profile depends on the DUT type and the relevant standard clause.

ISO 14708-7:2019 — Cochlear and Auditory Brainstem Implant Systems

Relevant Clause: 25.1 — Protection of the Active Implantable Medical Device from Damage Caused by Atmospheric Pressure Changes

This clause specifies low-pressure and high-pressure exposure conditions for evaluating cochlear and auditory brainstem implant systems.

Test Condition Standard Requirement
Low Pressure 50 kPa absolute
Low-Pressure Cycles 25 cycles
Low-Pressure Dwell Time Minimum 3 minutes
Low-Pressure Ramp-Up/Down Maximum 3 minutes each
High Pressure Minimum 304 kPa absolute
High-Pressure Cycles 40 cycles
High-Pressure Dwell Time Minimum 2 minutes
High-Pressure Ramp-Up/Down Maximum 2 minutes each

The standard requires the pressure values to be expressed as absolute pressure. The specimen is tested in saline solution (approximately 9 g/L), with the leads maintained at room temperature.

DUT evaluation requirements:

  • The pulse generator must provide uninterrupted pacing during pressure exposure.
  • After exposure, the pulse generator must operate as before the test without adjustment.
  • Permanent deformation is acceptable only when it does not adversely affect device operation, patient comfort or safety.

The KP3105 supports programmable low-pressure and high-pressure profiles, with pressure cycling and holding parameters configured according to the applicable test requirements.

ISO 14708-1:2014 — General Requirements for Active Implantable Medical Devices

Relevant Clause: 25.1

ISO 14708-1 Clause 25.1 addresses the protection of active implantable medical devices against damage caused by atmospheric pressure changes.

Pressure Condition Absolute Pressure Exposure Duration
Low Pressure 70 kPa ±3.5 kPa ≥1 hour
High Pressure 150 kPa ±7.5 kPa ≥1 hour

The evaluation considers the effects of deformation caused by the specified absolute-pressure conditions.

The KP3105 includes a 150 kPa ±7.5 kPa absolute-pressure holding capability for at least one hour. The 70 kPa low-pressure exposure profile can be included in the customized pressure-control configuration.

ISO 14708-2:2019 — Cardiac Pacemakers

Relevant Clauses: 25.1 and 25.2

ISO 14708-2 addresses particular requirements for active implantable cardiac pacemakers. Clause 25.1 refers to the general requirements specified in ISO 14708-1:2014 Clause 25.1.

Clause 25.2 specifies a water pressure chamber test to evaluate the resistance of implant system components to foreseeable pressure increases.

Test Condition Standard Requirement
Test Chamber Suitable water pressure chamber
Pressure Cycling 20 cycles
Pressure Transition Ambient pressure to maximum test pressure
Maximum Test Pressure 1.5 times the pressure specified in the manufacturer’s documentation
Pressure Change Rate ≥100 kPa/min
Maximum Pressure Holding ≥1 minute

KingPo KP3105 Cochlear Implant Atmospheric Pressure Cycling Test System with Liquid-Filled Chamber – ISO 14708-7:2019

The KP3105 provides a water pressure chamber with programmable pressure cycling and holding functions.

Its specified pressure change rate of ≥152 kPa/min, cycling capability of ≥40 cycles and holding time of ≥2 minutes provide margins above the minimum rate, cycle count and holding duration specified in Clause 25.2, provided that the required maximum pressure is within the configured equipment pressure range.

The final pressure target is determined according to the DUT manufacturer’s documentation.

Corresponding Chinese National Standards

The KP3105 can also be configured for the corresponding Chinese national standard requirements.

Chinese Standard Corresponding International Standard Main Scope
GB 16174.1-2024 ISO 14708-1:2014 (MOD) General requirements for active implantable medical devices
GB 16174.2-2024 ISO 14708-2:2019 (MOD) Particular requirements for cardiac pacemakers

The final test configuration should follow the applicable standard edition, relevant national modifications and the technical documentation of the DUT.

Test Principle and Workflow

The KP3105 uses a sealed water pressure chamber to establish a controlled pressure environment around the test specimen. The pressure-control system executes predefined pressure sequences, including pressure increase, pressure holding, pressure release and repeated cycling.

During testing, the pressure is adjusted according to the selected test profile. The system records pressure and time information to document the applied pressure conditions and the progress of the test.

Typical Testing Workflow

Specimen Preparation

Prepare the DUT according to the applicable standard and confirm the required specimen condition.

Specimen Installation

Position and secure the DUT using the appropriate fixture.

Parameter Configuration

Set pressure targets, cycle counts, ramp times and holding durations.

Pressure Application

Execute the programmed pressure increase, holding and release sequence.

Data Recording

KingPo KP3105 atmospheric pressure cycling test software interface for active implantable medical devices, featuring pressure test profiles, real-time pressure monitoring, cycling curves and gas circuit status.

Record pressure and time information throughout the test.

Post-Test Evaluation

Release chamber pressure, remove the specimen and evaluate its physical and functional condition.

The recorded pressure-time data can be used as supporting evidence for laboratory evaluation and test documentation.

DUT Installation and Specimen Preparation

The fixture arrangement should be selected according to the DUT dimensions, structure, mounting points and lead configuration.

Transparent pressure test chamber assembly of the KingPo KP3105 implantable medical device atmospheric pressure cycling system

For ISO 14708-7 Clause 25.1, the specimen must be tested in approximately 9 g/L saline solution, with leads at room temperature. The specimen immersion and containment arrangement must therefore be considered during system customization.

For ISO 14708-2 Clause 25.2, the DUT must be appropriately positioned in the water pressure chamber, with the maximum test pressure determined by the manufacturer’s documentation.

Typical Applications

The KP3105 is intended for pressure-related testing during the development, design verification and quality control of implantable medical devices, including:

  • Cochlear Implant Systems and auditory brainstem implant systems.
  • Active Implantable Cardiac Pacemakers and associated implant systems.
  • Other Active Implantable Medical Devices requiring atmospheric or water pressure resistance evaluation.

Typical users include third-party testing laboratories, medical device manufacturers, certification bodies and R&D laboratories.

KingPo KP3105 Atmospheric Pressure Cycling Test System

English technical datasheet with equipment specifications, ISO 14708 pressure test profiles, and specimen setup details.

Download Datasheet (PDF)

Compliance & Regulatory Assurance

The KingPo KP3105 is designed to support pressure-related testing requirements specified in the ISO 14708 series and corresponding Chinese national standards. The system can be configured according to the relevant test clauses and DUT-specific test procedures.

    • ISO 14708-1:2014, Clause 25.1 – Protection against atmospheric pressure changes.
    • ISO 14708-2:2019, Clauses 25.1 and 25.2 – Pressure-related protection requirements.
    • ISO 14708-7:2019, Clauses 25.1 and 25.2 – Atmospheric and water pressure-related test requirements.
    • GB 16174.1-2024 – Corresponding to ISO 14708-1:2014 (MOD).
    • GB 16174.2-2024 – Corresponding to ISO 14708-2:2019 (MOD).

Test pressure, pressure reference, cycling sequence, ramp rate and holding duration are determined by the applicable standard and DUT manufacturer’s specifications. The system configuration can be adapted to the required pressure profile and specimen arrangement.

Specimen conditioning and post-test functional evaluation are performed according to the applicable test procedure. For example, saline conditioning and functional assessment specified in ISO 14708-7 are specimen preparation and evaluation requirements, rather than assumed integrated functions of the pressure testing system.

Configuration and Technical Inquiry

The KP3105 is configured according to the applicable standard or DUT-specific test procedure. The pressure profile, chamber arrangement and specimen fixtures are determined by the required test conditions and product specifications.

Information Details
Applicable Standard Standard number, edition, relevant clause or specific test procedure.
DUT Specifications Device type, dimensions, product drawings and lead configuration.
Pressure Test Conditions Required pressure range, pressure reference, ramp rate, cycle count and holding time.
Specimen & Fixture Specimen dimensions, mounting arrangement and required test medium, such as water or saline solution.
Data & Reporting Pressure-time records, test data and preferred export or reporting requirements.

Please share the applicable standard or DUT technical documentation with KingPo. The system configuration can then be determined based on the actual testing requirements.

Detail Display

FAQ

What is the difference between atmospheric pressure testing and water pressure chamber testing?
Atmospheric pressure testing evaluates whether an active implantable medical device can withstand pressure changes during transportation and normal use. Water pressure chamber testing evaluates the resistance of implantable components to increased pressure under specified conditions. For example, ISO 14708-1 Clause 25.1 specifies low- and high-absolute-pressure exposure, while ISO 14708-7 Clause 25.2 specifies cyclic testing in a water pressure chamber. These are different test procedures and require different test configurations.
Is 304 kPa the required pressure for all active implantable medical devices?
No. The required pressure depends on the applicable standard and test clause. ISO 14708-2 Clause 25.1 specifies a minimum high-pressure condition of 304 kPa absolute for cardiac pacemaker testing. ISO 14708-1 Clause 25.1 specifies 70 kPa and 150 kPa absolute exposure, while ISO 14708-7 Clause 25.2 determines maximum pressure based on the manufacturer's specified pressure. The test pressure should therefore be selected according to the DUT documentation rather than applying 304 kPa universally.
Why is absolute pressure important in ISO 14708 pressure testing?
Absolute pressure is measured relative to a perfect vacuum, whereas gauge pressure is measured relative to local atmospheric pressure. ISO 14708 pressure test conditions specify absolute pressure values, so the pressure reference must be correctly configured and recorded to avoid deviations caused by atmospheric pressure differences.
Does active implantable medical device pressure testing require water or saline solution?
The test medium depends on the applicable test procedure. ISO 14708-2 Clause 25.1 specifies testing in saline solution at approximately 9 g/L, with the leads maintained at room temperature. ISO 14708-7 Clause 25.2 specifies a water pressure chamber test. Water and saline should not be treated as interchangeable test media. The chamber materials, seals, fixtures and specimen arrangement should be confirmed for the selected test method.
Can the KP3105 test both cardiac pacemakers and cochlear implants?
Yes, the KP3105 can be configured for relevant pressure testing of both device types. However, the test profiles and specimen arrangements are different. Cardiac pacemaker testing under ISO 14708-2 Clause 25.1 involves low- and high-pressure cycling in saline solution, while cochlear implant testing under ISO 14708-7 includes atmospheric pressure exposure and water pressure chamber testing.
Does the KP3105 monitor pacemaker operation during pressure exposure?
The KP3105's specified functions include pressure control and pressure-time data recording. Continuous pacing verification and post-test functional evaluation are separate requirements under ISO 14708-2. These checks require an appropriate monitoring or testing arrangement unless specifically integrated into the customized system configuration.
How is the maximum test pressure determined for ISO 14708-7 Clause 25.2?
The maximum test pressure is set at 1.5 times the pressure specified in the implantable device manufacturer's documentation. The test requires 20 pressure cycles, a pressure change rate of at least 100 kPa/min and a minimum holding time of one minute at maximum pressure.
What information is needed to configure an implantable medical device pressure testing system?
The most important information includes the applicable standard and clause, the manufacturer's specified maximum pressure, the required pressure reference, specimen dimensions, test medium and pressure cycling conditions. Product drawings are also useful for determining the chamber fixture and specimen positioning arrangement.

Related Products

Related News

KINGPO NEWS  ·  CMEF 2026  ·  POST-SHOW REPORT The 94th China International Medical Equipment Fair (CMEF) |  Beijing, China  |  […]

DONGGUAN, CHINA – October 20, 2026 – KingPo will participate in the 94th China International Medical Equipment Fair (CMEF), taking […]

External defibrillators are performance-critical medical devices. For third-party laboratories, manufacturers and engineering teams, a defibrillator tester—also commonly called a defibrillator […]

Shan Chao 1 , Qiang Xiaolong 2 , Liu jiming 3 , Zhang Chao 3 . (1. Heilongjiang Institute for […]

Global power supply specifications for households and industry: voltage, frequency, plug types, and standards for safe device use worldwide.

Abstract Thermage, a non-invasive radiofrequency (RF) skin tightening technology, is widely used in medical aesthetics. With operating frequencies increasing to […]

Technical Insight for IEC 60601-2-2 High-Frequency Surgical Equipment Testing High-frequency electrosurgical units (ESU) operate in the radio frequency range above […]

KingPo completed the delivery of a customized HME Moisture Loss Measurement System to UL Solutions Suzhou Laboratory on March 23, […]

Scroll to Top

Get A Free Quote Now !

Contact Form
If you have any questions, please do not hesitate to contact us.