KingPo TET thermal test setup showing transmitter and receiver test components, transparent water bath, robotic probe positioning and temperature sensor connections.
KingPo TET thermal test setup showing transmitter and receiver test components, transparent water bath, robotic probe positioning and temperature sensor connections.
KingPo transcutaneous energy transmission thermal testing system with KUKA six-axis robotic arm, transparent water bath, temperature measurement probes and integrated control cabinet.
KingPo robotic TET thermal testing equipment showing six-axis robotic positioning, transparent test tank, temperature probes and customized test fixtures.
KingPo transcutaneous energy transmission testing system with integrated touchscreen, temperature acquisition instruments, emergency stop and mobile equipment cabinet.
TP700 multi-channel temperature acquisition instrument integrated into the KingPo TET thermal testing system for PT1000 and NTC temperature sensor monitoring.

Transcutaneous Energy Transmission (TET) Testing System

The KingPo KP3100 TET Thermal Testing System evaluates local temperature rise in transcutaneous energy transmission assemblies for active implantable circulatory support devices, including ventricular assist devices (VADs). It integrates robotic positioning, a transparent test tank, multi-channel temperature acquisition and software for thermal data analysis.

  • Test Objects: TET transmitter and receiver assemblies
  • Test Purpose: Local temperature-rise and thermal behavior evaluation
  • Standard Reference: ISO 14708-5:2020, Clause 6.6.3.5.6
  • Robotic Positioning: KUKA six-axis robotic arm
  • Test Setup: Transparent tank with DUT-specific fixtures and probe arrangements
  • Temperature Acquisition: TP700 multi-channel temperature acquisition instrument
  • Data Analysis: Temperature-versus-time curves and temperature-rise (ΔT) calculation
  • Software Functions: Test case configuration, data recording, results review and report generation

Product Overview

Thermal Evaluation of Implantable Wireless Energy Transmission Systems

A Transcutaneous Energy Transmission System (TETS) transfers electrical energy from an external transmitter to an implanted device across intact skin, without requiring wires or tubes to penetrate the skin. This technology is used in certain active implantable medical devices, including circulatory support systems.

During TET operation, energy losses in the transmitting and receiving components may result in local heating. Temperature changes around these components and their surrounding test environment are therefore important considerations when evaluating the thermal performance of a TET system.

The KingPo KP3100 TET Thermal Testing System provides an integrated laboratory platform for this evaluation. A KUKA KR 4 R500-3 six-axis robotic arm establishes the required mechanical positioning of the test assembly, while a TP700 multi-channel temperature acquisition instrument records temperature changes at designated measurement locations.

PT1000 or NTC temperature probes can be arranged at selected positions to monitor temperature variations during TET operation. The integrated water bath provides a controlled test environment, while customized fixtures made from low-dielectric-constant materials support sample positioning.

The dedicated software supports test case configuration, temperature data acquisition, transient and steady-state curve display, temperature-rise calculation (ΔT), historical data review and test report generation.

The system is primarily intended for TET thermal evaluation, engineering research, product development and design verification associated with active implantable circulatory support devices. It focuses on local temperature measurement and does not replace the other electrical and functional measurements required for a complete TET performance evaluation.

Technical Specifications

The KP3100 integrates a six-axis robotic arm, multi-channel temperature acquisition equipment, a temperature-controlled water bath, customized test fixtures and dedicated software.

Robotic Positioning System

Parameter Specification
Robot Manufacturer KUKA
Robot Model KR 4 R500-3
Robot Configuration Six-axis articulated robot
Degrees of Freedom 6
Rated Payload 4 kg
Arm Reach 500 mm
Repeatability ±0.02 mm
Protection Rating IP54
Robot Body Weight 25 kg

Temperature Measurement System

Parameter Specification
Temperature Acquisition Instrument TP700 Multi-Channel Temperature Acquisition Instrument
Instrument Channel Capacity 1–200 channels
Channel Refresh Interval 1 s
Temperature Sensor Types PT1000 / NTC
Sensor Connection Three-wire or four-wire configuration
Specified Temperature Accuracy ±0.3°C

Water Bath and Test Fixtures

Parameter Specification
Water Bath Integrated temperature-controlled water bath
Temperature Stability ±1°C
Test Fixtures Customized fixtures
Fixture Material Low-dielectric-constant material

Test Software and Computer Requirements

Parameter Specification
Test Software Dedicated TET thermal testing software
Test Case Configuration Customizable according to test requirements
Data Processing Real-time monitoring, temperature curves and ΔT calculation
Report Generation Test report generation based on configured test cases
Operating System Windows 10 / 11
Processor Intel Core i5 or higher
RAM 8 GB or higher
Storage 100 GB or higher
Network Interface Gigabit Ethernet

Configuration Notes

The TP700 instrument specification supports 1–200 channels. The actual installed acquisition channels and temperature probes depend on the ordered configuration. The stated ±0.3°C temperature accuracy should be confirmed against the selected sensor and complete measurement chain.

System Components and Operation

The KP3100 combines mechanical positioning, temperature measurement, sample support and data processing in one laboratory platform. Each subsystem performs a specific function during TET thermal evaluation.

System Component Main Function Role in Testing
KUKA Six-Axis Robot Provides multi-axis mechanical positioning. Establishes the required DUT or test-assembly position within the configured working area.
Temperature-Controlled Water Bath Provides a controlled thermal environment for sample placement. Helps maintain the specified test environment during temperature-rise evaluation.
Customized Test Fixtures Secure and position the DUT and related components. Provide mechanical support while considering the dielectric properties of fixture materials.
PT1000 / NTC Probes Measure temperature at selected locations. Monitor local temperature variations around designated test positions.
TP700 Acquisition Instrument Collect multi-channel temperature signals. Transfers temperature measurement data to the test software for display and analysis.
Test Software and Computer Manage test cases and thermal measurement data. Displays temperature curves, calculates ΔT, stores historical records and generates reports.

System Operation

The DUT and its TET components are installed in customized fixtures and positioned within the test environment. The robotic arm establishes the required mechanical arrangement, while temperature probes are placed at designated measurement locations.

During operation of the TET system under test, the TP700 acquisition instrument collects temperature data from the configured channels. The dedicated software processes the recorded data, displays temperature variations over time and calculates temperature rise for subsequent engineering review.

The TET transmitter, implanted receiver and their operating power or control equipment are components of the DUT test setup. Their inclusion in the equipment supply scope should be confirmed separately.

Key Features

Six-Axis Robotic Positioning

The KUKA KR 4 R500-3 provides six-axis positioning for the TET test assembly. With a rated payload of 4 kg, an arm reach of 500 mm and specified repeatability of ±0.02 mm, it supports controlled positioning within the robot’s working area.

Multi-Point Temperature Measurement

The TP700 acquisition instrument supports PT1000 and NTC temperature probes with three-wire or four-wire connections. Multiple measurement points allow temperature changes to be monitored at selected locations around the TET components.

KingPo thermal testing software displaying multiple temperature measurement channels, real-time temperature curves and transient and steady-state thermal behavior during TET testing.

Customized Test Fixtures

Low-dielectric-constant fixture materials are used for DUT support and positioning. Fixture dimensions and mounting arrangements can be designed according to the transmitter, receiver and measurement requirements.

Temperature-Controlled Water Bath

The integrated water bath provides a controlled thermal environment for sample placement, with a specified temperature stability of ±1°C. The operating temperature range and test medium are determined by the test requirements.

Test Software and Data Analysis

The software supports test configuration, real-time temperature monitoring, temperature curve display, ΔT calculation, historical data review and test report generation.

Software Functions

The dedicated software provides a structured interface for configuring thermal test cases and reviewing temperature measurement results.

KingPo TET thermal testing software interface showing test case configuration, positioning distance, angular settings, temperature measurement points and real-time test monitoring.

Software Function Description
Test Case Configuration Create and configure test cases according to the DUT and selected thermal evaluation procedure.
Position Parameter Settings Provides test distance and tilt-angle setting fields for defining the test arrangement.
Temperature Measurement Setup Configure temperature measurement points and related probe parameters for the test case.
Real-Time Monitoring Display acquired temperature data during test operation for monitoring and review.
Temperature Curve Analysis Display transient and steady-state temperature curves to show temperature changes over time.
Temperature Rise Calculation Calculate temperature rise (ΔT) using the configured reference and measurement data.
Historical Data Review Search and review previously recorded test results.
Test Report Generation Generate test reports based on configured test cases and customer requirements.

Test Method and Workflow

Test Principle

The system evaluates the thermal behavior of a TET assembly by monitoring temperature changes at designated locations while the DUT operates under defined test conditions.

Temperature probes collect data from selected measurement positions. The TP700 acquisition instrument records the temperature signals, and the dedicated software displays the temperature response over time. Temperature rise (ΔT) is then calculated using the configured reference temperature and measurement data.

This approach enables engineers to review both the transient heating process and the subsequent steady-state temperature behavior of the TET assembly.

Typical Test Workflow

  1. Test Requirement Definition

    Determine the applicable test protocol, DUT configuration, operating conditions, required measurement positions and temperature evaluation criteria.

  2. DUT Installation

    Install the external transmitter, implanted receiver or relevant TET test components in the customized fixtures according to the planned test arrangement.

  3. Robotic Positioning

    Use the KUKA six-axis robotic arm to establish the required mechanical position. Set the relevant test distance and tilt angle according to the test plan and confirmed system configuration.

  4. Temperature Probe Arrangement

    Install PT1000 or NTC probes at the designated measurement locations and connect them to the configured TP700 acquisition channels.

  5. Test Environment Preparation

    Prepare the water bath and establish the required initial environmental conditions in accordance with the selected test procedure.

  6. TET Operation and Data Acquisition

    Operate the DUT using its specified TET power and control equipment. The system records temperature changes throughout the defined test period.

  7. Temperature Analysis

    Review transient and steady-state temperature curves, calculate ΔT and assess the results against the acceptance criteria defined in the applicable test protocol.

  8. Data Storage and Reporting

    Save the measurement data, review historical results and generate the corresponding test report.

Test Method Note

The actual test positions, sensor locations, test duration, DUT operating modes and acceptance criteria must be established by the applicable standard, device specifications and test protocol. Normal-operation or fault-condition testing must be configured through the DUT’s own operating setup where required.

KingPo Transcutaneous Energy Transmission thermal testing system in a laboratory, featuring a KUKA six-axis robotic arm, transparent water bath, temperature monitoring setup and integrated control cabinet for ISO 14708-5 related thermal evaluation.

Applicable Standards

ISO 14708-5:2020 – Circulatory Support Devices

Implants for surgery – Active implantable medical devices – Part 5: Circulatory support devices.

ISO 14708-5:2020 specifies requirements for the safety and performance of active implantable circulatory support devices, including ventricular assist devices (VADs), total artificial hearts (TAHs) and biventricular assist devices (biVADs).

Relevant Clause: 6.6.3.5.6 – Transcutaneous Energy Transmission Systems

This clause addresses TET systems that transfer electrical energy across intact skin to an implanted system without penetrating wires or tubes. It identifies technical characteristics that need to be specified and verified through appropriate analysis and testing.

Relevant TET evaluation parameters include:

  • Input power range
  • Output power range
  • Maximum power range
  • Energy transfer efficiency
  • Local temperature rise
  • Operational voltage range

The clause also addresses other TET-related performance and safety considerations. The KP3100 is specifically configured to support local temperature-rise evaluation through robotic positioning and multi-channel temperature measurement.

ISO 14708-1:2014 – General Requirements

Implants for surgery – Active implantable medical devices – Part 1: General requirements for safety, marking and for information to be provided by the manufacturer.

ISO 14708-1 establishes general safety requirements for active implantable medical devices. It provides the broader evaluation framework to be considered together with the particular requirements of ISO 14708-5 for circulatory support devices.

YY/T 0989.5-2022 – Chinese Medical Industry Standard

Implants for surgery – Active implantable medical devices – Part 5: Circulatory support devices.

This Chinese medical industry standard covers active implantable circulatory support devices, including ventricular assist devices, total artificial hearts and biventricular assist devices. It is relevant to TET-related thermal evaluation where the applicable Chinese test requirements and device-specific protocol call for such assessment.

Applications

VAD and LVAD Development

Thermal evaluation of transcutaneous energy transmission systems used in implantable ventricular assist devices, with temperature monitoring at designated locations during operation.

TET Subsystem Evaluation

Assessment of local temperature changes associated with external transmitters, implanted receivers and their relative positioning under defined operating conditions.

Medical Device Research and Testing

Support for thermal investigations, prototype evaluation and design verification in medical device R&D laboratories and third-party testing facilities.

Medical Device Manufacturing

Laboratory assessment of TET thermal performance during product development, engineering evaluation and design verification.

Application Note

For total artificial hearts (TAH), biventricular assist devices (biVAD) and other active implantable circulatory support devices, suitability depends on the use of a TET arrangement and compatibility with the configured test setup.

Compliance & Regulatory Assurance

The KingPo KP3100 is designed to support TET thermal evaluation in accordance with relevant requirements of ISO 14708-5:2020 Clause 6.6.3.5.6, ISO 14708-1:2014 and YY/T 0989.5-2022.

The system supports temperature monitoring and ΔT analysis for selected test requirements. Full device compliance requires evaluation against the applicable standards and test procedures.

KP3100 TET Thermal Testing System Datasheet

Explore the system configuration, temperature acquisition, software functions
and ISO 14708-5:2020 thermal evaluation scope.

Download Datasheet  ↓

Technical Inquiry & Expert Support

To determine the appropriate test configuration, please provide the following information:

Information Details to Provide
Standard & Test Method Applicable standard and clause, such as ISO 14708-5:2020 Clause 6.6.3.5.6, together with the relevant thermal testing procedure.
DUT & TET Configuration Device type (VAD/LVAD), transmitter and receiver dimensions, weight, arrangement and relevant technical drawings.
Positioning Requirements Required coil separation distance, alignment, tilt angle and robotic positioning conditions.
Temperature Measurement Required measurement locations, number of temperature points, sensor type (PT1000 / NTC) and acquisition channels.
Test Environment Required water bath temperature, test medium and sample placement conditions.
Data & Reporting Test duration, temperature recording requirements, data analysis and report format.

These details help KingPo evaluate the robotic positioning, fixture design, temperature measurement configuration and software requirements for your TET thermal testing application.

Detail Display

FAQ

What is a Transcutaneous Energy Transmission System (TETS), and why is thermal testing important?
A TETS transfers electrical energy from an external transmitter to an implanted receiver through intact skin. Thermal testing evaluates temperature changes around the transmitting and receiving components during operation, helping assess potential heating associated with energy transmission.
Which standard does the KingPo KP3100 TET Thermal Testing System support?
The KP3100 supports selected thermal evaluation requirements associated with ISO 14708-5:2020 Clause 6.6.3.5.6, which addresses transcutaneous energy transmission systems used in active implantable circulatory support devices.
What does the KP3100 measure during TET thermal testing?
The system monitors temperature changes at configured measurement locations, records temperature curves and calculates temperature rise (ΔT). It is primarily intended for local thermal evaluation rather than comprehensive electrical performance testing.
Which medical devices can be tested using the KP3100?
The system is primarily intended for TET-equipped ventricular assist devices (VADs), including left ventricular assist devices (LVADs). Other circulatory support systems, such as total artificial hearts (TAHs) and biVADs, may also be considered depending on their TET configuration and physical compatibility.
How does the six-axis robotic arm assist TET testing?
The KUKA KR 4 R500-3 robotic arm provides six-axis positioning for the test assembly. It supports controlled positioning according to configured test distances and angles, helping establish repeatable mechanical arrangements during thermal evaluation.
Which temperature sensors are supported by the KP3100?
The TP700 temperature acquisition instrument supports PT1000 and NTC temperature probes with three-wire or four-wire connections. The instrument supports 1–200 channels, while the actual installed channel capacity depends on the ordered configuration.
Can the KP3100 evaluate temperature changes at both the transmitter and receiver?
Yes. Temperature probes can be arranged at selected locations around the external transmitter, implanted receiver and other relevant test positions. The actual measurement points depend on the DUT structure, probe arrangement and available acquisition channels.
Does the KP3100 water bath simulate human tissue?
The system includes a temperature-controlled water bath with a specified temperature stability of ±1°C. However, the supplied configuration does not establish the water bath medium as validated tissue-equivalent material. The test medium should be selected according to the applicable test protocol.
Can the KP3100 measure TET energy transfer efficiency and output power?
The current configuration focuses on robotic positioning and temperature-rise measurement. Dedicated input power, output power, operating voltage and energy transfer efficiency measurement functions are not included. Additional measurement equipment is required if these parameters must also be evaluated.
What information is required to configure a KP3100 TET testing system?
Customers should provide the DUT type, dimensions and weight, transmitter and receiver specifications, applicable standard and test procedure, required positioning conditions, temperature measurement points and test environment. These details help determine the appropriate robotic positioning, fixture design, sensor arrangement and software configuration.