IEC 60601-1 7cm Strap Tester evaluates grip and handling device integrity, ensuring medical equipment safety and compliance.

  • Applies force uniformly over 7 cm strap length at handle center, ramping from 0 to test value in 5–10 s, holds 1 min.
  • Tests carrying handles/grips to withstand 4x equipment weight in any normal use/transport direction (Clause 9.4.4).
  • Force range 0–1500 N, time 0–9999 s, precise control for repeatable results.
  • Detects failure modes: loosening, permanent distortion, cracking, or breakdown.
  • Robust design with calibration for reliable IEC compliance in medical device Q

MOQ: 1
Price: To be quoted
Standard Packaging: safety carton pack or plywood box
Payment Method: T/T

Detail Information

Place of Origin China Brand Name Kingpo
Certification Iso9001 Ce Model Number KP-HD38-1
Standard: IEC60601-1-clause 9.4.4 Force: 0-1500N
Test Time: 0-9999s Handle Width: 7cm
Highlight: IEC 60601-1-Clause 9.4.4, 7cm Strap Tester, 1500n Strap Tester

Product Description

IEC 60601-1Grips and other handling devices-7 cm strap tester
Carrying handles or grips furnished on PORTABLE ME EQUIPMENT shall withstand loading as described in the following test:
The handles and their means of attachment are subjected to a force equal to four times the weight of the ME EQUIPMENT in any direction of NORMAL USE and transport.
If more than one handle is furnished on PORTABLE ME
EQUIPMENT , the force is distributed between the handles. The distribution of forces is determined by measuring the percentage
of the ME EQUIPMENT weight sustained by each handle with the ME
EQUIPMENT in the normal carrying position. If the ME EQUIPMENT is furnished with more than one handle but is so designed that it can readily be carried by only one handle, then each handle is to be
capable of sustaining the total force.
The force is applied uniformly over a 7 cm length of the handle at the centre, starting at zero and gradually increasing so that the test value will be attained in 5 s to 10 s and maintained for a period of 1 min.
Handles that break loose from the ME EQUIPMENT or exhibit any permanent distortion, cracking or other evidence of breakdown constitutes a failure.

TESTING AND MEASURING EQUIPMENT/ALLOWED SUBCONTRACTING

IEC 60601-1:2005 + Am.1:2012

Medical electrical equipment – Part 1: General requirements for basic safety and essential performance

“R” Required

“S” May be subcontracted, see OD 2012

“SPTL” Specialized Facility, see IECEE 02-2

“W” Witness testing in the categories “MED” and “MEAS”

“3PPS” Three Phase Power Supply required

Clause Measurement/testing Testing / measuring equipment / material needed Subcontracting
4.11 Power input Suitable devices for the voltage, current/power and frequency

Supply: 1 phase and 3 phase variacs

 

R
5.3 Ambient temperature, humidity, atmospheric pressure Suitable devices for recording ambient temperature, humidity, atmospheric pressure

 

R
5.7 Humidity preconditioning treatment Environmental conditions: Climate chamber controlling temperature and humidity

 

R
5.9.2 Accessible parts Force gauge (30 N), standard test finger (figure 6), straight unjointed test finger, test hook (figure 7)

 

R
7.1.2 Legibility of markings Illuminance meter

 

R
7.1.3 Durability of markings Distilled water, ethanol (96% pure), isopropyl alcohol , timer / stop watch

 

R
8.4.2 Accessible parts including applied parts Oscilloscope, oscilloscope leads, suitable instruments for measuring voltage, current, capacitance, test pin (Figure 8), metal test rod (D = 4 mm, L = 100 mm), force gauge (10 N)

 

R
8.4.3,

8.4.4

Limitations voltage and energy Suitable oscilloscope recorder /set-up & RCL meter R
8.5.5.1 Defibrillation protection 5 kV test circuit & oscilloscope interface circuit according to Figures 9 & 10, oscilloscope

 

S
8.5.5.2 Energy reduction test The test circuit according to Figure 11, oscilloscope, oscilloscope leads

 

S
8.6.4 Impedance and current-carrying capability Current source (25 A minimum, 50 or 60 Hz, 6 V maximum)

 

R
8.7 Leakage currents and patient auxiliary currents Measuring device according to Figure 12, mains isolation transformers, variacs, voltmeter, millivoltmeter, aluminium foil, diverse circuits (fig. 13-20)

 

R
8.8.3 Dielectric strength High voltage tester, isolating transformer for HV-tests (fig. 28), stop watch / timer

 

R
8.8.4.1 a) Ball pressure test Test equipment according to IEC 60695-10-2

 

R
8.8.4.2 Resistance to environmental stress Apparatus for ageing rubber in oxygen

 

S
8.9 Creepage distances and air clearances Oscilloscope, oscilloscope leads, callipers, micrometer, spacing gauges, force gauge (2 N & 30 N), standard test finger (figure 6)

 

R
8.9.1.7 Material groups classification Test equipment according to IEC 60112

 

S
8.9.3.4 Thermal cycling Heating cabinet

 

R
8.11.3.5 Cord anchorage Force gauge (at least 100 N), torque gauge (at least 0.35 Nm)

 

R
8.11.3.6 Cord guards Weights, angle gauge, radius gauge

 

R
9.4 Instability hazards 5° & 10° inclined planes or inclinometer or trigonometric calculation, force gauge (at least 220 N), 20 cm by 20 cm test surface, weights, test threshold (10 mm high and 80 mm wide), 7 cm strap, stop watch / timer

 

R
9.5.2 Cathode ray tubes The relevant tests of IEC 60065, Clause 18.

 

S
9.6.2.1 Audible acoustic energy A-weighted sound pressure level according to ISO 3746, ISO 9614-1 or IEC 61672-1

 

S
9.6.3 Hand-transmitted vibration Measurements are made in accordance with ISO 5349-1.

 

S
9.7.5 Pressure vessels Hydraulic pressure test apparatus

 

S
9.8 Hazards associated with support systems Weights or load cell, 0.1 m2 test surface, stop watch / timer, human body test mass (Figure 33)

 

R
10.1 X-radiation Radiation meter

 

S
10.3 Microwave radiation Radiation meter

 

S
10.4 Lasers Test equipment according to IEC 60825-1 S
11.1 Excessive temperatures Temperature indicator/recorder suitable for this function and

thermocouples, 4 wire resistance unit, test corner, variac

 

R
11.2 Fire prevention Spark ignition test apparatus (Figure 34), oxygen gas analyzer

 

S
11.3 Constructional requirements for fire enclosures FV tests specified in IEC 60695-11-10 S
11.6.2 Overflow 15° inclined plane or inclinometer or trigonometric calculation, stop watch / timer, high voltage tester

 

R
11.6.3 Spillage Flask or graduated cylinder, stop watch / timer

 

R
11.6.5 Ingress of water or particulate matter Classification tests of IEC 60529

 

W
11.6.6 Cleaning and disinfection Dielectric strength and leakage current tests as appropriate

 

R
11.6.7 Sterilization Sterilisation to client specification

 

S
13 Hazardous situations and fault conditions stop watch / timer, voltmeter, ammeter, temperature indicator / recorder suitable for this function and thermocouples, 4 wire resistance unit, cheesecloth

 

R
15.3 Mechanical strength Force gauge (250 N minimum), circular plane surface 30 mm in diameter, 500 g steel ball, 50 mm thick hardwood board (hardwood > 600 kg/m3), 40 mm step, hardwood doorframe (40 mm2), circulating air oven

 

R
15.4.2 Temperature and overload control devices Positive temperature coefficient devices (PTC’s) with IEC 60730-1: 1999, clauses 15, 17, J.15 and J.17

 

S
15.4.3.4 Primary Lithium batteries Performance of the tests identified in IEC 60086-4

 

S
15.4.3.4 Secondary Lithium batteries Performance of the tests identified in IEC 62133 S
15.4.6 Actuating parts of controls Force gauge (at least 100 N), torque gauge (at least 6 Nm), stop watch / timer

 

R
15.4.7 Cord-connected hand-held and foot-operated control devices Force gauge (minimum1 350 N), 30 mm diameter test tool, stop watch / timer

 

R
15.4.7.3 Entry of liquids Classification tests of IEC 60529

 

W
15.5.1.1 Transformers Winding tester for transformers, temperature indicator/recorder suitable for this function and thermocouples variac, loads

 

R
15.5.2 Dielectric strength 5x voltage / 5x frequency dielectric strength test fixture, stop watch / timer

 

R
16.6 Leakage currents As in sub-clause 8.7

 

R
A 10.4 Light emitting diodes (LEDs) As specified in IEC 62471

 

S
G Protection against hazards of ignition of flammable anaesthetic mixtures 4 mm & 12 mm diameter test rods, timer / stop watch, gasket test B-b of IEC 60068-2-2, force gauge (minimum 100 N), gas pressure meter (0 to 400 Pa range), voltmeter, ammeter, ohmmeter, meter for capacitance and inductance, vol/oxygen mixture (vol volume percentage 12,2 % ± 0,4 %),test apparatus for flammable mixtures (Figure G.7),

 

S
G.4.3 Prevention of electrostatic charges antistatic material testing according to ISO 2882

 

S
L Insulated winding wires for use without interleaved insulation Sample is preparation according to IEC 60851-5:1996 & dielectric according to 60601-1; flexibility and adherence test 8 of IEC 60851-3:1996; mandrels of diameters according to Table L.1; heat shock test 9 of IEC 60851-6:1996; 2 mm diameter shot of stainless steel, nickel or nickel plated iron. S

Get Free Quote of IEC 60601-1 Grips And Other Handling Devices-7 Cm Strap Tester

Ensure Handle Safety & IEC Compliance Choose the IEC 60601-1 7 Cm Strap Tester to benefit from:

  • Precise 7 cm uniform force application → Reliable Clause 9.4.4 results
  • Enhanced patient/operator safety → Prevent grip failures under load
  • Durable, calibrated design → Long-term accuracy in testing
  • Factory-direct support & fast delivery → Accelerate your projects
  • Cost-effective in-house verification → Reduce external lab expenses

Get Free Quote Now – Share your ME equipment weight/specs, and our team will propose the setup. Quick response, global shipping—contact KingPo today!

Applications

  • Testing carrying handles/grips on portable medical electrical equipment
  • Mechanical strength verification for patient transport devices
  • Compliance testing in medical device manufacturing and certification labs
  • Instability hazard assessment per Clause 9.4
  • QC for portable ventilators, monitors, infusion pumps, etc.
  • Safety validation during normal use and transport

FAQ

What is the IEC 60601-1 7 Cm Strap Tester used for?
It tests grips/handles on portable ME equipment to withstand 4x weight force without loosening or deformation (Clause 9.4.4).
Which standard does this tester comply with?
Fully compliant with IEC 60601-1 Clause 9.4.4 (2005 + A1:2012), includes calibration for reliable results.
How is force applied in the test?
Uniformly over 7 cm at handle center, ramps 5–10 s to test value, holds 1 min.
What is the MOQ for this strap tester?
MOQ is 1 unit, perfect for labs or compliance setup. Bulk orders qualify for discounts.
What force range does the tester support?
0–1500 N, suitable for most portable ME equipment (4x weight test).
Does it handle multi-handle designs?
Yes—force distributed based on normal carry position; single-handle tested at full force.
What failure modes does it detect?
Loosening, permanent distortion, cracking, or other breakdown indicating safety risk.
What warranty and support do you provide?
Factory support with calibration assistance and reliable after-sales service.
How does it improve medical device safety?
Ensures grips withstand transport/use loads, preventing instability hazards for patients/operators.
How does this tester reduce compliance costs?
In-house accurate testing avoids third-party fees, speeds certification, and ensures durable performance.

Related Products

Related News

Manufacturers perform glow wire tests by heating a wire and applying it to electrical components to assess fire resistance and meet safety standards.
KP-1065S Pulse Voltage Generator boosts surge testing efficiency, safety, and IEC 62368-1 compliance for advanced electronics and production lines.

For plastic insulation frames, given the high heat generated by the copper wires and iron core, if the flame retardant […]

ISO 5356-1:2015 is the internationally recognized standard titled Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and […]

IP (Ingress Protection) ratings are enclosure protection standards developed by the International Electrotechnical Commission (IEC), primarily assessing the ability of […]

Dynamic Compensation Implementation for High-Frequency Electrosurgical Unit Testing Using High-Frequency LCR or Network Analyzers Above MHz Shan Chao1, Qiang Xiaolong2, […]