Electrical Safety Test Equipment โ€“ KingPo IEC Testers

KingPo offers a comprehensive range of Electrical Safety Test Equipment designed to meet international standards such as IEC 60335, IEC 60601, IEC 60990 and more. Our precision instruments ensure reliable testing for leakage current, insulation resistance, dielectric strength, endurance, flexing, vibration, and special application safety compliance. Trusted by global laboratories, manufacturers in household appliances, medical equipment, electric tools, and personal care products.

IEC61215 Environmental Test Chamber Photovoltaic Modules Hail Impact Testing Machine

KingPo Main Electrical Safety Test Series

KingPo Electrical Safety Test Equipment includes specialized testers such as:

  • Flexing Test Apparatus (IEC 60335-1 & IEC 60335-2-23) for swivel and cord bending durability
  • Endurance & Force Test Machines (e.g. Toilet Seat Endurance Tester up to 1250N)
  • Leakage Current & Touch Current Testers (IEC 60990 Figures 3-5)
  • Vibration Measurement Systems for hand-held electric tools (IEC 60335 & ISO 28927-1)
  • Spark Ignition Testers for oxygen-enriched medical environments (IEC 60601-1)
  • Special application gauges and probes (e.g. durable test pins, lock security benches) All equipment is engineered for precision, repeatability, and compliance with global certification requirements.
Air Enthalpy Method Calorimeter Computer Control Interface โ€“ Real-Time Monitoring of Temperature, Pressure, Airflow, Power, EER/COP, Dehumidifying Capacity
IEC 60456 washing machine test chamber for measuring washing performance, water consumption, and energy efficiency of household washing machines and dishwashers.
Professional IEC 62552 Refrigerator Testing System for Part 3 Energy Consumption. ยฑ0.5โ„ƒ precision & 30-point matrix for TUV/VDE lab-ready certification data.
range hood airflow measurement test system with airflow duct and centrifugal fan
Vacuum Cleaner Air Performance Testing System Makes IEC 60312 Easy
IEC 61591 Air Volume Testing System โ€“ Air Chamber with Multiple Nozzles and Flow Equalizing Orifice Plate for Precise Range Hood Airflow Measurement
Hair Dryer Airflow Test System โ€“ High Precision Hot Wire Anemometer for Accurate Wind Speed Measurement
Dual Code Air Enthalpy Calorimeter Test Chamber with Indoor and Outdoor Sections for Air Conditioner Capacity Testing
Unsymmetrical Coupled Decoupling Network DATA-CDN-4B โ€“ KingPo Surge Test Equipment for 10/700ฮผs Voltage Wave up to 6 kV on 10/100/1000M Networks
IEC 60335-2-84 Test Chamber โ€“ KingPo KP-35841 with 0.055% Ammonia Precision for Toilet Seat Compliance Testing
KingPo KP-9571 Abrasion Resistance Test Machine and Hand-held Tool Vibration Measurement System.
KingPo IEC 60456 Water Supply System with integrated RO EDI and PLC control system.
KP1951S Pulse Generator
Plug Capacitor Discharge Tester
IEC 60335-2-80-Figure 102 Durable Steel Test pin
ANSI/BHMA A156.2 lock security test bench with motor-driven precision for bored locks. Grade 1-3 certification & ASTM F476 compliance.
1 2 3

Ready to ensure your products meet global safety standards?

Contact KingPo today for a free consultation, quotation, or technical discussion.
๐Ÿ“ฉ Email:[email protected]
๐Ÿ“ž Tel: +86-769-81627526

Electrical Safety Test Equipment FAQs

What standards do KingPo electrical safety testers comply with?

Primarily IEC 60335 series, IEC 60601-1, IEC 60990, ISO 28927-1, EN 60745, etc.

What is the latest version of IEC 60335-1 and what are the key 2025โ€“2026 changes?

IEC 60335-1:2020 (Edition 6.0) remains the current base standard. Between 2024 and 2026, most Part 2 standards have been aligned to Edition 6, with important updates including:
โ€ข Mandatory use of Test Probe 18 and Probe 19 for accessibility
โ€ข Stricter external surface temperature limits
โ€ข Enhanced requirements for remote operation and cybersecurity (Annex U, Clauses 22.40, 22.51)
โ€ข Strengthened battery-operated appliance rules (Annex B, Clause 12)
โ€ข Updated rules for appliances using flammable refrigerants (IEC 60335-2-24, -2-89, -2-40:2024)
KingPo test equipment has been updated to support all 2020 Edition 6 clauses and the latest Part 2 alignments (e.g. IEC 60335-2-9:2023/A11:2023 enforcement Jan 2026). Contact us for a free gap analysis.

How is leakage current tested under IEC 60335 and what are the limits?

Leakage current is measured according to Clauses 13 and 16 using the measuring network of IEC 60990 Figure 4. Typical limits (at operating temperature):

  • Portable/hand-held appliances: โ‰ค 0.75 mA (normal condition), โ‰ค 1.5 mA (single fault)
  • Stationary/fixed appliances: โ‰ค 3.5 mA (normal), โ‰ค 5 mA (single fault) KingPo leakage current testers provide high-precision RMS measurement (resolution ยฑ5 ยตA, bandwidth up to 1 kHz) and are fully compatible with IEC 60990. Most common failure: insulation contamination or insufficient creepage/clearance leading to re-test or recall.
What are the dielectric strength (hipot) test requirements in IEC 60335?

Clause 16 requires a withstand voltage test immediately after the temperature-rise test:

  • Basic insulation (Class I): 1 250 V
  • Supplementary insulation (Class II): 2 750 V Test duration: 1 minute, short-circuit current > 5 mA (IEC 61180-1). For routine production-line testing (Annex A): 1 000โ€“2 500 V, 1โ€“60 seconds. KingPo hipot testers support 0โ€“5 kV AC/DC with automatic current limiting (5โ€“30 mA range) and meet CBTL calibration requirements.
What are the new battery safety requirements for cordless appliances in IEC 60335?

Edition 6 significantly strengthened Annex B (battery-operated appliances) and Clause 12:

  • Charging, over-charge, short-circuit, and abnormal operation tests
  • Metal-ion battery specific requirements (UN 38.3 T.3/T.4 vibration & shock referenced)
  • Leakage current limit often tightened to โ‰ค 0.5 mA in many cordless products KingPo battery endurance testers, force applicators, and charging simulation fixtures help verify compliance for cordless vacuum cleaners, power tools, and IoT devices. Brazil INMETRO enforcement for many battery products is scheduled for 2026.
What are the most common reasons products fail IEC 60335 certification?

According to data from TรœV, Intertek, Eurofins, and CB scheme reports, the top failure modes are:

  1. Leakage current exceeding limits (most frequent โ€“ ~50% of failures)
  2. Excessive touchable surface temperature (Clause 11)
  3. Insufficient creepage/clearance distances (Clause 29)
  4. Abnormal operation leading to fire/electric shock hazard (Clause 19 โ€“ PEC + EMI double fault)
  5. Mechanical strength / accessibility failures (Clause 21 โ€“ Probes 11, 13, 18, 19) KingPo full suite of IEC 60335 testers (leakage, hipot, temperature rise, flexing, vibration, abnormal operation) + pre-compliance checklist helps catch issues early in R&D and avoid expensive re-testing.
How do I test smart / connected appliances for remote operation and cybersecurity under IEC 60335?

Edition 6 introduced Annex U (software evaluation) and new clauses (22.40, 22.51) requiring:

  • Validation of remote control safety
  • Simulation of software faults and network attacks
  • Abnormal operation under PEC + EMI double fault conditions KingPo abnormal operation and endurance testers can simulate fault scenarios, followed by leakage current re-check after EMC exposure. This is a growing hotspot due to the rapid increase in IoT household appliances.
What are Test Probes 18 and 19 and when must they be used?

Probe 18 (small unjointed test finger) and Probe 19 (very fine probe) were introduced/strengthened in Edition 6 alignments for better simulation of child / small-object access:

  • Probe 18: used to check accessibility to live parts or hot surfaces
  • Probe 19: finer probe for gaps in grilles, vents, etc. KingPo supplies complete IEC 61032 / IEC 60335 test probe kits (Probes 11, 13, 18, 19, etc.) with calibrated force applicators (10 N, 20 N, 30 N) to meet the latest 2025โ€“2026 requirements.
What are the new requirements for appliances using flammable refrigerants (R290, etc.)?

IEC 60335-2-24, -2-89, -2-40:2024 editions allow higher charge limits in some cases but demand:

  • Refrigerant detection systems
  • Ignition risk assessment (spark ignition test)
  • Leakage simulation under Clause 19
  • Double-wall heat exchangers resistant to freezing KingPo spark ignition testers, endurance machines, and leakage simulation apparatus are fully compliant with these updated safety clauses.
What equipment is needed for 100% production-line testing under Annex A of IEC 60335-1?

Annex A routine tests typically include:

  • Protective earthing continuity (โ‰ค 0.1 ฮฉ at 10 A)
  • Dielectric strength (1 000โ€“2 500 V, 1โ€“60 s)
  • Functional / leakage current check KingPo integrated production-line testers combine hipot, ground bond, and leakage current in one station with barcode traceability and ISO 17025 calibration.
Does KingPo IEC 60335 test equipment support the latest 2026 standards? What is the lead time?

Yes โ€“ KingPo equipment supports IEC 60335-1:2020 Edition 6 + over 100 Part 2 standards (including -2-7 washing machines, -2-24 refrigerators, -2-40 heat pumps, -2-89 commercial appliances, etc.).

  • Standard models: 15โ€“30 working days
  • Customized fixtures / probes: 25โ€“45 days Free technical consultation and gap analysis available to ensure your products meet 2026 enforcement dates (CB, CE, INMETRO, etc.).
Scroll to Top

Get A Free Quote Now !

Contact Form
If you have any questions, please do not hesitate to contact us.