Standards
IEC 60601-1
IEC 60601-2-2
zuozhe

IEC 60601-2-2:2017+AMD1:2023 TESTING AND MEASURING EQUIPMENT

TESTING AND MEASURING EQUIPMENT/ALLOWED SUBCONTRACTING

IEC 60601-2-2:2017+AMD1:2023, Edition 6.1

Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

IEC 60601-2-2
IEC 60601-2-2

Download IEC 60601-2-2-2017 standard

Clause Measurement/testing Testing / measuring equipment / material needed Equipment
201.8.4.101 Neutral electrode monitoring circuit Test setup as per Figure 201.103, sound level meter
201.8.4.102 Neuromuscular stimulation Capacitance meter,
dc resistance meter

KP8850

ESU Analyzer

201.8.5.1.2 MOPP Caliper,
high voltage tester
201.8.5.2.3 Patient leads Caliper,
high voltage tester
201.8.7.3.101 Thermal effect of HF leakage current RF ammeter,
load and measuring resistors,
tape measure, setup according to figures 201.104 – 201.107
201.8.8.3.101 Active accessory insulation saline solution,
caliper,
timer
201.8.8.3.102 Active accessory HF leakage saline solution,
caliper,
HF voltage source,
RF ammeter,
HF capacitance meter,
tape measure
201.8.8.3.103 Active accessory HF dielectric strength saline solution,
HF voltage source ,
caliper,
porous cloth, metal foil,
timer
201.8.8.3.104 Active accessory mains frequency dielectric strength saline solution,
HF voltage source ,
caliper,
porous cloth, metal foil,
timer
201.8.10.4.2 Connection cords Cord anchorage test apparatus as per Figure 201.108, aperture, weights, tape measure, dc source, ammeter
201.8.10.4.101.4 Footswitch Force gauge (force ≥ 10N) applied over an area of 625 mm2
201.11.6.3 Spillage Measuring cup (1liter), timer, high voltage tester
201.11.6.5 Ingress of water or particulate matter saline solution, caliper, timer, impedance analyser or other suitable devices for measuring ac impedance
201.11.8 Interruption of the power supply Electrosurgical analyser or other suitable devices to measure output power
201.12.1.101 Accuracy of output control setting Electrosurgical analyser or other suitable devices to measure output power, oscilloscope, various load resistances, setup as per Figures 201.109 and 201.110
201.12.1.102 Monotonicity of output control setting Electrosurgical analyser or other suitable devices to measure output power, oscilloscope, various resistances, setup as per Figures 201.109 and 201.110
201.12.1.103 Accuracy of maximum output voltage Oscilloscope
201.12.4.2 Indication of parameters relevant to safety Electrosurgical analyser or other suitable devices to measure output power
201.12.4.2.101 Output indicator Frequency analyser, sound level meter, tape measure
201.12.4.3.101 Output reduction means Electrosurgical analyser or other suitable devices to measure output power
201.12.4.4.101 Maximum allowed output power in single fault condition Electrosurgical analyser or other suitable devices to measure output power
201.12.4.4.102 Output power during simultaneous activation Electrosurgical analyser, RF ammeter or other suitable devices to measure output power and HF current, test setup as per Figure 201.111
201.13.2.13.101 Protection against the effects of short-circuiting of the electrodes Timer
201.15.4.1.102 Retention of detachable active electrodes Pull tester, timer
201.15.101.2 NE cord attachment DC or mains frequency, ammeter
201.15.101.3 NE cord connector, no conductive parts on patient Test finger as per Figure 6 of general standard
201.15.101.4 NE cord insulation Tests as per 201.8.8.3.102a), 201.8.8.3.103, 201.8.8.3.104
201.15.101.5 NE thermal performance HF voltage source,
RF Ammeter,
temperature scanning apparatus,
caliper,
environmental chamber (23±2°C), human subjects or surrogate medium or test device as defined in this clause,
timer
201.15.101.6 NE contact impedance HF voltage source,

20cm x 30 cm flat metallic plate, true r.m.s.,
suitable true r.m.s. a.c. ammeter,
HF LCR meter

201.15.101.7 NE adhesion Timer, pull tester, caliper, 0.9% saline solution

 

 